CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 4.1 Rack Sensor Check
3 - 61
3
When E Sensor is blocked, any other sensors blocked will cause a Flashing LED for that
sensor.
Example: B blocked and E blocked = LED II Flashing
B blocked, E NOT blocked, = LED II ON.
7. Verify E Sensor is ON just before B Sensor.
a. Place normal rack at D Sensor (
b. Ensure LED IV is ON.
c. Slowly move another normal rack into position at B Sensor (
) while observing
LEDs II and IV.
d. Verify LED IV flashes just before LED II.
“VP - 4.2 Rack Sensor Adjustment” (Page 3-63)
.