VERIFICATION PROCEDURESVP - 4.1 Rack Sensor Check
3 - 59
3
VP - 4.1 Rack Sensor Check
Purpose: Perform this procedure to verify Rack Sensor Operation.
All Loaders
1. Remove all Racks from Sample Loader.
2. Clean all Racks according to Operator’s Manual.
3. Clean Rack Sensors A through E with cotton tip applicator and water (
).
4. Ensure sensor surface does not protrude beyond platen wall. Sensor face must be flush or
recessed not more than 1/2 thickness of platen wall (
). If sensor position is incorrect,
go to
“VP - 4.2 Rack Sensor Adjustment” (Page 3-63)
5. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
c. Set BAR (1/0) Switch to 1.
Special Tools
DESCRIPTION
QTY
PN
Feeler Gauge Set
1
1-43627-01