CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 4.1 Rack Sensor Check
3 - 60
3
Figure 3-27 Rack Sensor Side View
6. Check Rack Sensors.
a. Ensure Platen is clear of loose items and all ALARM LEDs are OFF.
b. Put a 0.100" (2.5mm) shim and Normal Rack (not End Rack) at positions shown
c. Push rack tight against shim.
d. Verify ALARM LED is ON for each position (See
).
Platen
Wall
1/2 Thickness
Of Wall
Sensor
Must NOT
Protrude Beyond
Platen Wall
SL-156A