Warnings

• This device was sterilized with ethylene oxide and is for single use only. Do not reuse or resterilize this device. Attempts to 

resterilize this device can cause a malfunction, insufficient sterilization, or harm to the patient.

• Do not use this device if the sterile package is open or damaged.
• The delivery sheath is designed to be used with the loader. Do not attach a syringe directly to the delivery sheath because 

the luer is incompatible and may result in ingress of air or excessive bleeding.

• Use the hemostasis valve to impede the backflow of blood during the implant procedure.
• Do not use a power injection syringe to inject contrast solution through the delivery sheath.

Precautions

• This device should be used only by physicians who are trained in standard transcatheter techniques. The physician should 

determine which patients are candidates for procedures that use this device.

• The physician should exercise clinical judgment in situations that involve the use of anticoagulants or antiplatelet drugs 

before, during, and/or after the use of this exchange system. 

• Use on or before the last day of the expiration month that is printed on the product packaging label. 
• Use caution when advancing the sheath and dilator to avoid damaging tissue and vessels or interfering with previously 

implanted medical devices.

• Store in a dry place.

Potential Adverse Events

Potential adverse events that may occur during or after a procedure using this exchange system may include, but are not limited 
to:

Procedure

CAUTION: When placing a device using an AMPLATZER™ TorqVue™ Exchange System, refer to the instructions for use 
provided with the device.
General instructions for the AMPLATZER™ TorqVue™ Exchange Systems are provided below.

 1.  Prepare the exchange system for use:

- Inspect the exchange system sterile pouch and verify that it is unopened and undamaged. Do not use the exchange 

system if the sterile barrier has been compromised.

- Gently open the sterile pouch and inspect the components for damage. Do not use damaged or kinked components.
- Flush all components with sterile saline.
- Wipe the delivery sheath with sterile gauze moistened with sterile saline to remove any foreign material.

2. If the plastic vise is attached to the original delivery cable, remove the plastic vise.
3. Screw the distal end of the exchange system delivery cable to the proximal end of the original delivery cable.
4. Insert the dilator into the delivery sheath. You may encounter resistance as the dilator reaches the distal end of the 

sheath because the last few inches of the sheath are tapered. Turn the rotating luer on the dilator clockwise to lock the 
components together.

5. Remove the original delivery sheath from the patient.
6. Advance the exchange system delivery sheath and dilator assembly over the delivery cable until the delivery sheath is 

positioned according to the device’s instructions for use
CAUTION: If the distal tip of the sheath needs to be advanced into the left atrium of the heart, advance the delivery 
sheath and dilator assembly only as far as the vena cava before removing the dilator and purging air from the delivery 
sheath. 

7. Remove the dilator and allow back-bleeding to purge any air from the delivery sheath. 

• Air embolism

• Infection

• Arrhythmia

• Myocardial infarction

• Arteriovenous fistulae

• Perforation

• Bleeding at the access site

• Peripheral pulse loss

• Brachial plexus injury

• Stroke

• Cardiac tamponade

• Thrombosis

• Death

• Valve damage

• Dissection

• Vascular occlusion

• Endocarditis

• Vessel damage

• Hematoma

Содержание Amplatzer TorqVue

Страница 1: ...ath Provides a pathway through which a device is delivered B Loader Introduces a device into the delivery sheath C Hemostasis valve with extension tube and stopcock Allows flushing of the exchange system and controls back bleeding D Delivery cable Attaches to an in vivo delivery cable to facilitate removal of a delivery sheath from a patient Attaches to a device for controlling its movement throug...

Страница 2: ...ral instructions for the AMPLATZER TorqVue Exchange Systems are provided below 1 Prepare the exchange system for use Inspect the exchange system sterile pouch and verify that it is unopened and undamaged Do not use the exchange system if the sterile barrier has been compromised Gently open the sterile pouch and inspect the components for damage Do not use damaged or kinked components Flush all com...

Страница 3: ...zard waste procedures Warranty Abbott Medical warrants to buyer that for a period equal to the validated shelf life of the product this product shall meet the product specifications established by the manufacturer when used in accordance with the manufacturer s instructions for use and shall be free from defects in materials and workmanship Abbott Medical s obligation under this warranty is limite...

Страница 4: ...ence number Product serial number Batch Code Use by date Use on or before the last day of the expiration month noted on the product packaging Do not reuse Sterilized using ethylene oxide Unique device identification Consult instructions for use Follow instructions for use on this website Keep dry Do not use if package is damaged Not made with natural rubber latex Inner diameter Outer diameter Leng...

Страница 5: ...ting Recommended delivery sheath catheter dimensions Federal law USA restricts this device to sale by or on the order of a physician or properly licensed practitioner Quantity Date of manufacture Exchange System ONLY Exchange System ...

Отзывы: