2.2.5 Standard RTD probe installation
The standard RTD measures flowing gas temperature and should only be used if
the G4 EX is located in a General Purpose area, following local codes. The
procedures that are presented in this section will enable the user to install the
RTD into the meter run.
2.2.5.1
Optionally supplied materials
-
RTD probe with 10’ of cable; optional lengths are 15’, 25’, 30’ 40’ and 50’
-
One (1) Thermowell
with ¾” npt threads; optional threads are ½” and 1”
-
Nylon tie wraps
2.2.5.2
Materials not supplied
-
Customer must specify Thermowell
“U” length
-
Teflon
®
tape
2.2.5.3
Step-by-step instructions
1) Install the Thermowell into the meter run (see Figure 2
The power should be removed from the G4 EX before
performing any field wiring.
2) Using snap ring pliers, adjust the probe length so that it is spring loaded
against the bottom of the Thermowell.
3) Screw the probe connector into the Thermowell.
Figure 2
–5 Thermowell installation into the meter run
4) At the G4 EX, remove the plug from the explosion proof RTD hub, and
screw the adapter bushing into the hub.
5) Remove the nut from the water tight cord connector. Leave the nylon
sealing ring in place.