2
IN/RandC/007-EN Rev. A
| SM500F | Paperless recorder | Battery replacement
4 Replacing the battery
1.
Isolate the recorder from the power supply.
Referring to Fig. 4.1:
2.
Unlock the door with the key (if required)
A
, press release
catch
B
and open the door.
Referring to Fig. 4.2:
3.
Use a T8 Torx driver to remove 6 Torx screws
C
securing
the door assembly rear cover and remove the rear cover.
Referring to Fig. 4.3:
4.
Disconnect processor board flexi-cable
D
, display
flexi-cable
E
keypad membrane cable
F
and inverter
cable (if fitted)
G
from the processor board. Lift out and
discard the processor board.
5.
Turn the processor board over.
Referring to Fig. 4.4:
6.
Remove battery
H
and insert replacement.
Referring to Fig. 4.3:
7.
Position the processor board in the door housing and
reconnect inverter cable
G
(color recorders only – if fitted)
keypad membrane cable
F
, display flexi-cable
E
and
processor board flexi-cable
D
.
Referring to Fig. 4.2:
8.
Refit the door assembly rear cover and use a T8 Torx driver
to secure 6 Torx screws
C
.
9.
Close and lock the door.
10. Restore the recorder's power supply.
CAUTION – Property damage
The recorder is vulnerable to electrostatic damage.
Wear an anti-static strap or place the instrument on
an anti-static workbench.
Fig. 4.1 Opening the door
Fig. 4.2 Removing the rear cover
A
B
C
Fig. 4.3 Disconnecting / removing the processor board
Fig. 4.4 Replacing the battery
D
E
F
H