Navigator 500
Sodium analyzer
3 Installation
12
OI/ASO550–EN Rev. G
3.2 Installing the transmitter
3.2.1 Transmitter optional accessories
Optional accessories comprise:
Cable gland kit
3.2.2 Transmitter location
For transmitter general location requirements refer to Fig. 3.6.
Install in a clean, dry, well ventilated and vibration-free location
providing easy access. Avoid rooms containing corrosive gases
or vapors, for example, chlorination equipment or chlorine gas
cylinders.
3.2.3 Panel mounting
Warning.
The transmitter is not fitted with a switch – an
isolation device such as a switch or circuit breaker
conforming to local safety standards must be fitted to the
final installation. It must be fitted in close proximity to the
transmitter, within easy reach of the operator and marked
clearly as the isolation device for the transmitter.
Fig. 3.6 Transmitter location
Eye-level location
Humidity
Ambient temperature
55 °C
(131 °F)
Max.
–10 °C
(14 °F)
Min.
0 to 95 %
Avoid vibration
Dimensions in mm (in.)
* To DIN43700
**
150 mm (6 in.) if (optional) cable glands fitted
Fig. 3.7 Transmitter panel-mount option
92
(3.6)
35
(1.4)
Panel-mount kit –
optional
Cable gland kit
– optional
Mounting panel –
maximum thickness
6 (0.236)
Panel cut-out dimensions*
40 (1.57)
40 (1.57)**
186
(7.32 )
+0.04
–0
+1.1
–0
186
(7.32 )
+0.04
–0
+1.1
–0
Panel clamp anchor screw
torque values: 0.5 to 0.6 Nm
(4.42 to 5.31 lbf/in.)