TB82PH ATEX Product Instruction Manual
ABB
16
TB82PH-ATEX-EN-A
CALIBRATE MODE
Before an instrument can properly measure pH the sensor must be calibrated to the instrument. Periodic update
calibrations must be performed during the time the sensor is in service. The frequency of re-calibration is application
dependent. This mode also defines calibration of the temperature input and fine tuning the 4-20 mA output for non-
fieldbus versions.
MEASURE
CALIBRATE
pH Cal
1 Point Calibration
Stable?
Enter New Value
2 Point Calibration
Change Temp
O
C
Low Value Buffer
Stable?
High Value Buffer
Stable?
PV Slope %
PV Offset mV
Temperature Cal
Enter Temp
Edit Calibration
Enter Slope
Enter Offset
Reset Cal
Output Cal
Adjust 4 mA
Adjust 20 mA
Back to pH Cal
OUTPUT/HOLD MODE
MEASURE
CALIBRATE
OUT/HOLD
Hold Output
Release Hold
Hold to XXX%
Rerange Output
4 mA Point
20 mA Point
Damping
XX.X Seconds
Back to Hold Output
Output/Hold screen flow only valid for non-fieldbus versions.
Caution.
The Reset Calibrate State will reset all
calibration values; therefore, the process sensor
and temperature sensor will require calibration
after performing the Reset Calibrate procedure.