User Guide IM/MM2WM–EN Rev. A

MagMaster

Electromagnetic flowmeter

MagMaster to WaterMaster 
electronics upgrade

Introduction

These instructions describe how to upgrade an existing 
MagMaster MF/F flowmeter (DN80 to DN1600) 
installation with remote transmitter for use with the new 
remote WaterMaster transmitter.

This guide should be read in conjunction with the 
following publications:

– WaterMaster FET100 User Guide, OI/FET100–EN

– WaterMaster Programming Guide, IM/WMP

SP033 Adam Street Wynnum SPS - WaterMaster - Electromagnetic Flowmeter - 

OIML R49 Class 1 450mm - User Guide

Q-Pulse Id: VM486

Active: 08/04/2016

Page 1 of 12

Содержание MagMaster MF/F

Страница 1: ...DN80 to DN1600 installation with remote transmitter for use with the new remote WaterMaster transmitter This guide should be read in conjunction with the following publications WaterMaster FET100 User Guide OI FET100 EN WaterMaster Programming Guide IM WMP SP033 Adam Street Wynnum SPS WaterMaster Electromagnetic Flowmeter OIML R49 Class 1 450mm User Guide Q Pulse Id VM486 Active 08 04 2016 Page 1 ...

Страница 2: ...d unrivalled service and support The quality accuracy and performance of the Company s products result from over 100 years experience combined with a continuous program of innovative design and development to incorporate the latest technology Quality Control The UKAS Calibration Laboratory No 0255 is just one of the ten flow calibration plants operated by the Company and is indicative of our dedic...

Страница 3: ... Cable 3 2 1 MagMaster Sensor Cable Wiring to WaterMaster Transmitter 3 2 1 1 Sensor Coil Resistance 3 3 Programming Configuration 4 3 1 Accessing the Programming Menus 4 3 2 Programming Menu Sensor Setup 4 SP033 Adam Street Wynnum SPS WaterMaster Electromagnetic Flowmeter OIML R49 Class 1 450mm User Guide Q Pulse Id VM486 Active 08 04 2016 Page 3 of 12 ...

Страница 4: ...ofit units set the DIP switches as shown in Fig 1 2 Leave SW1 SW2 unaltered as they are not part of the retrofit upgrade detailed here Note These upgrade instructions and the use of WaterMaster with existing MagMaster sensors is supported only on the MF F product code style of sensors not the MF E type Fig 1 1 Configuration DIP Switches Fig 1 2 Configuration DIP Switches Note The backplane fitted ...

Страница 5: ...ith the meter turned off Fig 2 1 Sensor Cable Connections at Transmitter Terminal Block Standard System Caution Do NOT reverse the inner signal E1 E2 and outer screen S1 S2 cores of the coaxial cables Ensure the black anti microphonic layer in the grey and white screened cables is removed Note An alarm on the WaterMaster display indicates if the coil resistance is too low for the WaterMaster trans...

Страница 6: ...is menu disappears If this menu is not present and you are unable to configure the necessary calibration factors below contact ABB Service The Sensor Setup menu while visible enables the setting of items such as the Sensor Type Sensor Size and calibration Span Ss and Zero Sz at Advanced level access These items are normally editable only at Service level and are hidden after setup Sensor Setup Sen...

Страница 7: ...fied on the data label identified as Cal 1 It is also available on the meter calibration certificate 2 Multiply the MagMaster Cal Fact 1 by the Cal Fact 1 to Ss Multiplier to find the Span Ss value 3 Enter the correct Span Ss value as calculated 4 Ensure flow is present in the sensor 5 If the flow reads in the wrong direction enter the span as a negative value for example 123 45 instead of 123 45 ...

Страница 8: ...3 1 0 1 350 104 9 0 1 400 104 5 0 1 450 104 0 0 1 500 105 0 0 1 600 105 4 0 1 700 105 4 0 1 760 101 4 0 1 800 101 2 0 1 900 101 9 0 1 1000 101 8 0 1 1050 101 8 0 1 1100 101 8 0 1 1200 101 8 0 1 1400 101 8 0 1 1500 101 8 0 1 1600 101 8 0 1 Table 3 1 WaterMaster Span Ss Values SP033 Adam Street Wynnum SPS WaterMaster Electromagnetic Flowmeter OIML R49 Class 1 450mm User Guide Q Pulse Id VM486 Active...

Страница 9: ...ed the WaterMaster transmitter reboots automatically After reboot re login and continue setup Coil Measurement F Frequency to use for sensor coil diagnostic measurements Dependent on sensor size Ensure the Drive Mode is set first Refer to Table 3 2 for the correct value to enter Sensor Size Drive Mode Coil Measurement F DN350 5A 40 Hz DN350 5A 20 Hz DN400 5A 20 Hz DN450 5A 20 Hz DN500 5A 20 Hz DN6...

Страница 10: ...libration correction factor Set to Zero for retro fit sensors 0 000 Sc Calibration correction factor Set to Zero for retro fit sensors 0 000 Cal Status Status of calibration settings for this flowmeter Only when programming of calibration settings is complete fully checked and the flowmeter is working correctly adjust this setting to Calibrated The Sensor Setup Not Complete alarm should now disapp...

Страница 11: ... and Sensors Ammonia Nitrate Phosphate Silica Sodium Chloride Fluoride Dissolved Oxygen and Hydrazine Analyzers Zirconia Oxygen Analyzers Katharometers Hydrogen Purity and Purge gas Monitors Thermal Conductivity Customer support We provide a comprehensive after sales service via a Worldwide Service Organization Contact one of the following offices for details on your nearest Service and Repair Cen...

Страница 12: ... purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document We reserve all rights in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior w...

Отзывы: