FAULT FINDING
Refer to Fig. 7.1 for fault diagnosis procedure. If a fault is found which cannot be resolved using the procedure detailed in Fig. 7.1,
return the AquaProbe Transmitter to The Company.
Check
transmitter
and cabling
Check
transmitter
and cabling
Adust VU, TU
or PU
Turn on
flow
Install
AquaProbe
correctly and/or
ensure correct
functioning
Connect
AquaProbe
cable
Reading
reversed ?
AquaProbe
correctly
installed &
functional
?
Flow in
Pipeline?
AquaProbe
cable
connected?
Display
response correct?
See Fig. 6.1
Correct
flowrate
displayed?
Totaliser
operating
correctly?
Flow in
forward direction
& count rate
< 10Hz?
Check
transmitter
and cabling
Replace
battery
Display
permanently
blank?
Ensure power
is correct and
connector is
correctly wired
Remove and
reconnect
power
Display
blank?
End
Reset parameters for
correct flow direction
& count rate at <10Hz
Connect power
to unit
Start
Display
?
Yes
No
No
No
No
Yes
Yes
Yes
Yes
No
No
Yes
Yes
Yes
Yes
No
No
Yes
Yes
No
No
No
LO P
Fig. 7.1 Fault Diagnosis