Access via GSM

Commissioning test for signal strength

This provides a means of carrying out a radio
signal test for selecting the optimum location for
the antenna. The system can be tested in its final
commissioned location and state (eg manhole
cover closed and any local comms equipment
disconnected from the meter, etc, etc).

Menu 12 controls the GSM features.

Go to Menu 12 and select the 

Signal Test Wait

Time

 [>354].  Enter here the time in seconds that

the system will wait before starting the signal test.
A count down starts from the selected Wait time to
zero.  This count down is shown on the LCD
Display.  At this point the installation should be
closed up into its commissioned state.  When the
count reaches zero, a radio signal strength
measurement is taken and the result is shown on
the Display for 30 seconds (long enough to open
the door or meter cover to inspect the result).

The highest signal strength is represented by a
value of 31; the poorest strength is a value
below 5.

SIM Numbers

SIMs have both a Voice Telephone Number and a
Data Telephone Number.  Use the Voice Number
for Remote Access via SMS Text;  use the Data
Number for Remote Dial Up operation e.g. using
LogMaster.

GSM WakeUp Control

Battery Powered Meters

To conserve battery power, the GSM radio module
is normally powered down. To operate remote
communications (SMS Requests, Auto Report or
Remote Dial Up), the GSM radio module can be
woken either by going to menu 12 and requesting
a Manual Wakeup [>358=1] - which will force the
module to wake up for the WakeUp Duration
[>352] - or by setting a programmable WakeUp
schedule as shown below. The Schedule [>353]
can be programmed for wake ups every 12 hours,
1 day, 7 days or  Always Off. The Base Time [>351]
sets the time of day for the wake ups and the Base
day [>347] sets the day of the week (for the 7 day
schedule).

WIRELESS COMMUNICATIONS

Mains Powered Meters

When the meter is mains powered, the GSM radio
module is continuously powered and is able to
handle remote communications (SMS Requests,
Auto Report or Remote Dial Up). If mains power is
interrupted and the meter is powered from the
back up battery, the GSM module will not operate.

Note.

 On mains powered GSM units the

module is automatically rebooted at midday
every day.  This process takes approx 1 minute
and it is therefore recommended that the Base
Time should be set to avoid the period around
midday.

WakeUp Duration

The power supply to the transmitter may have
been ordered as one of the following types :

Mains with Battery Backup

Battery only (dual internal cells)

External Battery Pack with Backup

For Mains powered units with battery backup, the
WakeUp Duration does not apply as the GSM
module is either continuously powered (whilst
there is mains power present) or not powered

(whilst on backup battery).

For Battery only units, powered either by 2 internal
cells or by an external battery pack, the WakeUp
Duration (>352) has a range of 3 to 23 

minutes

.

Schedule [>353]

ON

Time

Base Time [>351]
Base Day [>347]

ON

OFF

On Time [>352]

Содержание FieldIT AquaMaster

Страница 1: ...FieldIT Electronic Commercial Water Meter with Integral Logger AquaMasterTM Quick Reference Guide IM AMG QRG_1 ...

Страница 2: ...oted 1 The relevant sections of these instructions must be read carefully before proceeding 2 Warning labels on containers and packages must be observed 3 Installation operation maintenance and servicing must only be carried out by suitably trained personnel and in accordance with the information given 4 Normal safety precautions must be taken to avoid the possibility of an accident occurring when...

Страница 3: ...ther versions of Windows vary From the Windows Start menu choose Programs Accessories Hyperterminal HyperTerminal At New connection enter AquaMaster At Connect using choose COM1 or COM2 depending on the PC connection At Port setting prompt choose Bits per second 4800 Data bits 8 Parity None Stop bits 1 Flow control none Programming the AquaMaster To access programming mode connect the AquaMaster t...

Страница 4: ...formation AquaMaster ABB Limited Stonehouse etc Tab Esc Tab Note Passwords are case sensitive and may be changed by the user in menu 4 0 PROGRAMMING Further key operations access the menus as follows Key Menu Displayed Text User entered data Function Key Note Pressing M within a menu displays all the remaining parameters immediately There are two levels of password control Level Default Menus Pass...

Страница 5: ...e for no error 42 is the current flow rate 3 is the error code for Write Access Denied since writing to the flow rate is not allowed To change the flow FSD to 10 m3 min To cancel the password security press Esc until the front screen is reached Changing Selections Note Ensure that the correct access level is selected It is possible to enter data directly into the AquaMaster without navigating the ...

Страница 6: ...enu system has been activated by pressing 3 or 4 times and that a suitable login level has been obtained go to a given Menu by typing its number followed by Tab eg to go to menu 12 type 12 and AquaMaster menu 12 will now be accessible 12 0 GSM Comms Settings PARAMETER CHANGES Tab Tab ...

Страница 7: ... D t t a B t f e L R 2 3 2 s y a D s n i a M t t a B t h g i R R 5 4 2 s y a D t t a B t f e L v e r P R 9 9 2 s y a D t t a B t h g i R v e r P R 4 3 2 m h o k A g i S R 5 3 2 m h o k B g i S R t i x E 0 c e s m 1 o t y t i c o l e v w o l f e h t s t e s e d o M t s e T v e R d w F s t c e f f a y l n O s e s o p r u p t s e t r o f t u p t u O e s l u P d n a s r e s i l a t o t t e N Z y l n o...

Страница 8: ... l a G l m l D G M m3 h l t f 3 s l a G D L M t f 3 m t f 3 m3 s t f 3 h l M m3 m s l a g U m3 h m l a g U m3 d h l a g U s l a G D G U M s e t o N e e s s g n i t t e S e r u s s e r P 0 6 6 6 e d o M 2 9 1 1 s t i n U e r u s s e r P 2 2 2 1 V R U r o 0 0 1 D S F 2 3 2 1 V R L r o 0 o r e Z 2 0 2 1 r a B r e p s t i n U l a i c e p S 2 1 2 1 e m a N e r u s s e r P l a i c e p S 2 t i x E 0 s t ...

Страница 9: ... e m i T t c e n n o C l a t o T 4 4 5 3 s e m i T t i a W t s e T l a n g i S 4 8 5 3 n o i s s e S M S G l a u n a M 4 s g n i t t e S p u e k a W c i d o i r e P 1 5 3 e m i T e s a B p U e k a W 4 7 4 3 y a D e s a B p U e k a W 4 3 5 3 e l u d e h c S p U e k a W 4 2 5 3 n o i t a r u D p U e k a W 4 t i x E 0 s e c i v r e S S M S 0 3 1 2 6 3 o N l e T S M S 4 t r o p e R o t u A S M S 1 6 3...

Страница 10: ...module can be woken either by going to menu 12 and requesting a Manual Wakeup 358 1 which will force the module to wake up for the WakeUp Duration 352 or by setting a programmable WakeUp schedule as shown below The Schedule 353 can be programmed for wake ups every 12 hours 1 day 7 days or Always Off The Base Time 351 sets the time of day for the wake ups and the Base day 347 sets the day of the we...

Страница 11: ...a mobile phone to AquaMaster and an SMS Reply Text with the requested information will be sent to the originating phone or SMS gateway If the AquaMaster is battery powered the SMS Request will not be serviced until the next programmed WakeUp time as defined in Menu 13 WIRELESS COMMUNICATIONS SMS Request Message SMS Request messages must have the following format password command command command pa...

Страница 12: ...t the GSM module is not powered up the timestamp of the SMS message containing the report may be significantly different from that on which the report was created Adding the TIM command to the Auto Report command string 363 will include in the report the time date that the report was created Care should therefore be taken when setting the wakeup time day and the AutoReport schedule to avoid confli...

Страница 13: ...PIN Setting AquaMaster to use SIM s existing PIN Simply change SIM Password to be the PIN of the SIM card AquaMaster will automatically use this to send to the GSM module each time it powers the module up If this is incorrect the SIM will eventually become locked out Using AquaMaster to change SIM s PIN 1 Change SIM Password using the format oldpin newpin 2 The new PIN can be seen in SIM Password ...

Страница 14: ... d e m i t y a D t h g i n d n a y a d r e t n i w g n i r u d e m i t t h g i N r e m m u s g n i r u d 3 s e m i t y a d l l A s e m i t t h g i n l l A 4 r e m m u s g n i r u d e m i t t h g i N t h g i n d n a y a d r e m m u s g n i r u d e m i t y a D r e t n i w g n i r u d 5 r e m m u s g n i r u d t h g i n d n a y a D r e t n i w g n i r u d t h g i n d n a y a D 6 g n i r u d e m i t t...

Страница 15: ...13 NOTES ...

Страница 16: ...14 NOTES ...

Страница 17: ...15 NOTES ...

Страница 18: ...16 NOTES ...

Страница 19: ... pH conductivity and dissolved oxygen transmitters and sensors ammonia nitrate phosphate silica sodium chloride fluoride dissolved oxygen and hydrazine analyzers Zirconia oxygen analyzers katharometers hydrogen purity and purge gas monitors thermal conductivity Customer Support We provide a comprehensive after sales service via a Worldwide Service Organization Contact one of the following offices ...

Страница 20: ...ined herein without notice Printed in UK 03 04 ABB 2004 ABB Inc 125 E County Line Road Warminster PA 18974 USA Tel 1 215 674 6000 Fax 1 215 674 7183 ABB Limited Oldends Lane Stonehouse Gloucestershire GL10 3TA UK Tel 44 0 1453 826661 Fax 44 0 1453 829671 ABB has Sales Customer Support expertise in over 100 countries worldwide www abb com ...

Отзывы: