…5 Configuration

Sensor Setup

Menu

Sensor Setup

Exit

Select

Used to set the tag, measurement units, operational range and clean functions and to 
compensate for salinity and barometric pressure.

Menu

Comment

Default

S1 (to 4) :DO 

Select the ID of the optical dissolved oxygen probe to be cleaned and scroll to the 
Clean Interval menu to configure the cleaning routine.

 

Tag
PV Type
Units
Range High
Range Low
Filter Type

These parameters are not used to configure cleaning – refer to the probe Operating 

instruction (OI/ADS420-EN) for parameter descriptions.

Clean Interval

Set the interval between cleans:  

Off / 15 Mins. / 30 Mins. / 45 Mins. / 1 to 24 Hours

Off

Clean Type *

Set the clean type: Continuous / Pulsed

Continuous

Clean On Time *

Set the duration of the clean: 1 to 60 Secs

30 Secs

Clean Off Time * / **

Set the duration between cleans: 1 to 60 Secs

30 Secs

Number of Pulses * / **

Set the number of cleaning pulses: 1 to 10 Pulses

2 Pulses

Recovery Time *

Set the time delay between the completion of cleaning and the display of a new 

reading on the operator page: 1 to 10 Min.

1 Min.

Clean Duration *

Displays the total duration of the clean: 
Clean Type set to Continuous = Clean on Time + Recovery Time

Clean Type set to Pulsed = (Clean on Time + Clean Off Time) 
x Number of Recovery Time

Clean Output *

Displays the output signal the clean is assigned to. This can be set to relay 1 to 6 or 
digital output 1 to 6 –  refer to AWT420 transmitter Operating instruction 

 

(OI/AWT420-EN).

No Assignment

Salinity Correction
Barometric Pressure
Restore Defaults

These parameters are not used to configure cleaning – refer to the probe Operating 

instruction (OI/ADS420-EN) for parameter descriptions.

Salinity Units 

Pressure Units 

* Displayed only if Clean Interval is NOT set to Off

** Displayed only if Clean Type is set to Pulsed

Содержание EZClean

Страница 1: ... transmitter to enable automatic scheduled cleaning of the probe For more information Publications for the associated AWT420 transmitter and ADS420 optical dissolved oxygen probe are available for free download from www abb com measurement or by scanning these codes AWT420 transmitter ADS420 probe Search for or click on Commissioning Instruction AWT420 Universal 4 wire dual input transmitter CI AW...

Страница 2: ...unting 7 Pole mounting 8 4 Connections 9 Pneumatic connections 9 Connect the tubing to compressed air unit 9 Connect the tubing to the ABB compressed air adaptor 9 Relay connections 10 Mains voltage supply 10 5 Configuration 11 Sensor Setup 12 6 Maintenance 13 Annual maintenance 13 7 Specifications 14 8 Spares and accessories 15 Spares 15 EZClean compressor maintenance kit 15 Air cleaning tubing k...

Страница 3: ...ed air supply The EZClean system generates compressed air up to a pressure of 3 5 bar g 50 7 psi g maximum do not use compressed air for any other purpose than that for which it is provided never direct a stream of compressed air towards your body or the body of any other person when handling connecting compressed air lines the correct safety procedures must be observed suitable PPE must be worn g...

Страница 4: ...those involved in the life cycle of electrical and electronic equipment In conformity with European local and national regulations electrical equipment marked with the above symbol may not be disposed of in European public disposal systems after 12 August 2005 NOTICE For return for recycling please contact the equipment manufacturer or supplier for instructions on how to return end of life equipme...

Страница 5: ...lean component descriptions Item Description Item Description A Compressor motor H Air out bulkhead push fit coupling to 6 mm nylon tubing B Compressor head I M20 cable gland to transmitter C Air muffler filter J M20 cable gland mains supply voltage D Air out to sensor K Air intake including dust filter E Pressure relief valve L Instrument label F Mains supply voltage connector M Door locks G Fuse...

Страница 6: ... sufficient space in front of the unit to open the door fully 280 mm 11 0 in Dimensions in mm in 330 13 0 395 15 5 205 8 1 381 15 0 262 10 3 420 16 5 440 17 3 196 7 7 Add 280 11 0 when door open Figure 2 Wall mounting Referring to Figure 2 1 Fit the four mounting lugs A to the compressor unit B in the required plane and secure using the M6 screws washers C 2 Mark appropriate mounting lug screw hol...

Страница 7: ...2 1 6 U bolt 210 8 3 21 0 8 420 16 5 385 10 1 262 10 3 170 6 70 195 7 67 486 19 1 70 2 75 Figure 3 Handrail mounting Referring to Figure 3 1 Fit the four handrail mounting plates A to the compressor unit B in the required plane and secure each plate with one M6 screw C supplied If optional extension mounting plates D are required attach to the handrail mounting plate s in the required plane using ...

Страница 8: ...1 330 13 0 50 1 9 U bolt 205 8 1 25 1 0 42 1 6 U bolt 205 8 1 21 0 8 420 16 5 Add 280 11 0 when door open Figure 4 Pole mounting Referring to Figure 4 1 Fit the pole mounting plate A to the compressor unit B and secure using the M6 screws C supplied 2 Position the compressor unit B on the pole and secure using the U bolt D and M8 nuts washers E supplied NOTICE It is not usually necessary to fit a ...

Страница 9: ...onnect one end of the 6 mm OD nylon tubing A to bulkhead push fit connector B Figure 5 Connecting 6 mm OD nylon tubing to the compressed air unit 2 Proceed to Connect the tubing to the ABB compressed air adaptor to connect 6 mm OD nylon tubing to ABB compressed air adaptor Connect the tubing to the ABB compressed air adaptor Referring to Figure 6 1 Refer to the sensor accessories instructions for ...

Страница 10: ...L1 connection on compressed air unit Figure 7 Relay connections compressed air unit to AWT420 transmitter 4 Tighten the cable glands on the compressed air unit and transmitter 5 Proceed to Mains voltage supply to make the mains voltage supply connection to the compressed air unit Mains voltage supply 230 or 115 V AC WARNING Isolate the compressed air unit and AWT420 transmitter from the mains powe...

Страница 11: ...sociated probe as follows Set Type to Output Set Source to Sx Clean where x corresponds to the sensor position 1 to 2 at the EZLink connector 5 Exit the Input Output menu and use the keys to scroll to the Sensor Setup menu Menu Sensor Setup Exit Select Press the key the Sensor Setup page is displayed 1 2 3 Sensor Setup S1 DO Salinity Units Pressure Units Back Select If more than 1 sensor is connec...

Страница 12: ...ecs 30 Secs Clean Off Time Set the duration between cleans 1 to 60 Secs 30 Secs Number of Pulses Set the number of cleaning pulses 1 to 10 Pulses 2 Pulses Recovery Time Set the time delay between the completion of cleaning and the display of a new reading on the operator page 1 to 10 Min 1 Min Clean Duration Displays the total duration of the clean Clean Type set to Continuous Clean on Time Recove...

Страница 13: ...umatic connections on page 9 3 Ensure the air intake is clean see System overview on page 5 item K 4 Ensure the air muffler filter is clean and free of clogged pores see System overview on page 5 item C To clean this remove from the compressor head and blow oil free compressed air through it from the reverse side If the unit operates in a dusty environment additional periodic cleaning may be requi...

Страница 14: ...n 2 0 bar g 30 psi g Typical operating pressure and flow rate 1 38 bar g 20 psi g 14 16 L min 0 5 scfm set to perform every 4 hours TORR BAR 0 200 400 600 760 0 5 1 0 1 5 2 0 25 20 15 10 5 30 0 10 20 30 10 20 0 1 0 3 0 4 0 5 0 6 0 7 0 8 0 9 0 2 INS HG PSIG LITRES M SCFM Figure 9 Pump unit pneumatic pressure flow rate Environmental ratings Storage temperature 20 to 50 C 4 to 122 F Operating tempera...

Страница 15: ...pressor pneumatic fittings kit Part number Description 3KXA494400L0042 Compressor pneumatic fittings kit Accessories EZClean compressor pole mounting kit Part number Description 3KXA494400L0017 Compressor pole mounting kit includes compressor pole mounting plate U bolts nuts and washers EZClean compressor handrail mounting kit Part number Description 3KXA494400L0018 Compressor handrail mounting ki...

Страница 16: ...ossible lack of information in this document We reserve all rights in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior written consent of ABB Copyright ABB 2022 All rights reserved 3KXA494406R4201 ABB Measurement Analytics For your local ABB contact...

Отзывы: