EL6010 Gas Analyzer in Category 2G Operator’s Manual
42/24-30 EN Rev. 1
Chapter 4
Start the Gas Analyzer
CAUTION!
If the EL6010-... analyzer unit must be opened for operation or maintenance,
the instructions on the following warning label (affixed to the housing) must
be observed:
“
Before opening, disconnect power and wait 10 minutes!
”
For operation of the EL6010-CU control unit the following warning instruc-
tion must be observed:
“
Do not open when an explosive atmosphere may
be present!
”
Check the Installation
Check the Installation
Make sure the EL6010 gas analyzer is correctly installed before carrying out any
start-up procedures.
Use the following check list:
Check
9
Installation site
•
Do the conditions at the installation site (zone rating, explosion group,
temperature class) match the information on the identification plate?
•
Are the control unit and the analyzer unit not installed outdoors?
•
Are the control unit and the analyzer unit securely fastened?
Operation with sample gases at positive pressures
•
Is the device version designed for operation with sample gases at
positive pressures? The identification plate should read: “See Special
Conditions for Sample Gas Pressure”.
Connection of gas lines
•
Are all gas lines correctly connected?
•
Is the pressure sensor connection not joined with the sample gas path?
Potential compensation connection
•
Is the external potential compensation connection of the analyzer unit
connected to the local potential compensation point?
•
Is the external potential compensation connection of the control unit
connected to the local potential compensation point?
Continued on next page