CORIOLISMASTER FCX100, FCX400
SEALING OF DEVICES | IN/FCX100/FCX400/MID/OIML-XA REV. A
EN – 13
7. Continue to twist the seal wire
1
up to the front head
hole screw
2
. Secure the front head hole screw
2
as
described previously.
G12362
3
4
1
2
1
Fig. 22
8. Continue to twist the seal wire
1
to the left eyelet
2
on the lower cover. Guide one end of the wire through
the eyelets on the cover and housing, guide the other
end on the outer side near the eyelets. Tightly twist the
seal wire
1
right after the left eyelet
2
.
9. Continue to twist the seal wire
1
up to the right
eyelet
3
. Secure the right eyelet
3
as described
previously.
10. Tightly twist both ends of the seal wire after the right
eyelet
3
over a length of approx. 5 cm.
11. Place the seal
4
approx. 5 cm from the last eyelet
3
onto the ends of the seal wire and press together using
sealing pliers. Cut off any excess
1
seal wire.
3.6
FCT400 transmitter with single-compartment
housing
NOTICE
Before the device is sealed, any settings must be
completed and the write protection switch must be set to
“write protection active” (see chapter ‘Activating the
hardware write protection’ on page 3 ).
Scope of delivery
Designation Quantity
Seal 1
Seal wire
G12363
3
1
4
2
1
2
Fig. 23
1. Insert the seal wire
2
through the openings on the
upper eyelet
1
such that both ends are the same
length.
2. Tightly twist the seal wire
2
right after the upper
eyelet
1
.
3. Continue to twist the seal wire
2
up to the lower
eyelet
3
. Both ends of the seal wire must be very close
together and the twist loops should be smaller than 1
cm.
4. Secure the lower eyelet
3
as described previously.
5. Tightly twist both ends of the seal wire after the lower
eyelet
3
over a length of approx. 5 cm.
6. Place the seal
4
approx. 5 cm from the lower eyelet
3
onto the ends of the seal wire and press together using
sealing pliers. Cut off any excess
1
seal wire.