Connecting the CoreSense M10
53
—
Figure 34
Mains AC power cable configuration
LINE
NEUTrAL
EArTh
hU Live
hU Neutral
Earth
AC Live
100–240 V AC, single phase
50–60 hz
8A max.
Copper-only wire
(stranded WITh FErrULE or solid)
AWG #18 to AWG #12
90 °C 600V, UL and CSA type
Flat screwdriver
Ø 6.0 mm
AC Neutral
AC Earth
Flat screwdriver
Ø 3.5 mm
NOTICE
The mains power cable must be able to withstand a minimal top operating temperature
of 90 °C.
Connecting the cable
NOTICE
All terminals are properly identified in the instrument itself to simplify installation.
To connect the cable:
1
Insert the cable through the power entry port and route it under the cable guide found on the left-
hand side of the analytical unit cabinet (see Figure 35 on page 53).
—
Figure 35
Routing the mains power cord toward the power entry terminal block
Power
entry port