C1300
Advanced Circular Chart Recorder
7 Modbus Registers
IM/C1300–MOD
Rev. C
19
7.5 Totalizer Settings
Register
Number
Read/Write
Description
Response/Entry
191
192
193
194
195
196
197
198
R
R
R
R
R
R
R
R
Predetermined Value Channel 1*
High Word
Low Word
Preset Value Channel 1*
High Word
Low Word
Front-panel Totalizer Value Ch. 1*
High Word
Low Word
Secure Totalizer Value Ch. 1*
High Word
Low Word
The Limit for High Word:Low Word linked
together is 0 to 99,999,999
201
202
203
204
205
206
207
208
R
R
R
R
R
R
R
R
Predetermined Value Ch. 2*
High Word
Low Word
Preset Value Channel 2*
High Word
Low Word
Front-panel Totalizer Value Ch. 2*
High Word
Low Word
Secure Totalizer Value Ch. 2*
High Word
Low Word
The Limit for High Word:Low Word linked
together is 0 to 99,999,999
211
212
213
214
215
216
217
218
R
R
R
R
R
R
R
R
Predetermined Value Channel 3*
High Word
Low Word
Preset Value Channel 3
*
High Word
Low Word
Front-panel Totalizer Value Ch. 3*
High Word
Low Word
Secure Totalizer Value Channel 3*
High Word
Low Word
The Limit for High Word:Low Word linked
together is 0 to 99,999,999
221
222
223
224
225
226
227
228
R
R
R
R
R
R
R
R
Predetermined Value Channel 4*
High Word
Low Word
Preset Value Channel 4*
High Word
Low Word
Front-panel Totalizer Value Ch. 4*
High Word
Low Word
Secure Totalizer Value Channel 4*
High Word
Low Word
The Limit for High Word:Low Word linked
together is 0 to 99,999,999
*These values are a combination of the High Word and Low Word.