Aztec 600 ISE ammonia and fluoride
Single-stream ion-selective analyzers
3 Installation
12
OI/AXM630–EN Rev. I
3.4.2 Mounting the Analyzer
Referring to Fig. 3.3:
1. Mark the wall using the dimensions shown.
2. Drill and plug 3 holes
A
and
B
, suitable for M6 or
1
/
4
in.
fixings.
3. Screw in top fixing
A
, leaving a gap of 20 mm
(0.78 in.) between the fixing head and the wall.
4. Hang the analyzer onto fixing
A
, ensuring the analyzer is
retained firmly against the wall.
5. Secure the analyzer to the wall using 2 fixings
B
.
3.4.3 Mounting the Optional Reagent Tray
If used, place the reagent tray no more than 1100 mm (43.3 in.)
from the analyzer’s bottom tray – see Fig. 3.2, page 11.
Referring to Fig. 3.4:
1. Mark the wall using the dimensions shown.
Alternatively, support the tray against the wall and mark
through the mounting holes.
2. Drill and plug mounting holes suitable for M8 or
5
/
16
in.
fixings.
3. Secure the tray to the wall using M8 or
5
/
16
in. fixings.
Dimensions in mm (in.)
Fig. 3.3 Mounting the Analyzer
Note.
Clearance – the enclosure doors can open 180°. If
mounting in a confined area, allow sufficient clearance for
cables on the door hinge side (min. 270 mm [10.6 in.]) and
100 mm (3.93 in.) on door opening side.
Note.
It is not possible to adjust fixing
A
once the
analyzer is placed over it. If necessary, remove the
analyzer and adjust the fixing.
460
(18.1)
150 (5.9)
Dimensions in mm (in.)
Fig. 3.4 Reagent Tray (Optional)
565 (22.2)
600 (23.6)
235
(9.2)
300 (11.8)
10
(0.39)
300 (11.8)
6 Holes Ø
8.5 (0.34)
35
(1.4)
35
(1.4)
400
(15.7)