Aztec 600 aluminium, ammonia, color, iron, manganese, phosphate
Single- and multi-stream colorimetric analyzers
12 Diagnostic Information and Icons
56
IM/AZT6CR–EN Rev. R
12 Diagnostic Information and Icons
12.1 Analyzer Diagnostic Information
The analyzer is programmed to display diagnostic messages to
provide information on servicing requirements and any other
conditions that develop during operation.
All diagnostic messages displayed on the analyzer are added to
the analyzer's audit log.
Table 12.1 shows icon types, diagnostic messages and ON/OFF
status for the relays.
Note.
The diagnostic icons in the following table conform to
NAMUR 107.
Diagnostic Icon
NAMUR Status
Failure
Check function
Out of specification
Maintenance required
Icon
Diagnostic message
Stop
Relay
Attention
Relay
Calibration
Relay
Failure
Relay
Monitor in Service
The analyzer is measuring correctly.
OFF
OFF
OFF
OFF
20 Character User-defined Alarm Tag
User-defined alarm message.
–
–
–
–
Monitor Stopped
Analyzer operation has been stopped by the user.
ON
OFF
OFF
OFF
Monitor Off
Analyzer operation has not been started or a failure has caused the
analyzer to stop operation and switch off all services.
ON
OFF
OFF
OFF*
Calibrating
A calibration routine is being run.
OFF
OFF
ON
OFF
A/D Error
A hardware error exists on the main board.
Power the analyzer down, wait for 10 seconds and power-up again.
If the error still exists contact the local ABB representative
OFF
OFF
OFF
ON
*Relay status shows ON if a failure has caused the stoppage.
Table 12.1 Diagnostic Information, Relay Operations and Inhibits (Sheet 1 of 5)