ABB AZ200 737 Скачать руководство пользователя страница 6

6

IN/ANAINST/032–EN Rev. B

| Endura AZ20 series probe | Combustion oxygen monitor

6 Refitting the cell and large surface 
area filter

Referring to Fig. 6.1, page 7:
1.

Check the cell mounting area for damage and ensure it is 
clean and dry. Clean by hand only using a non-metallic pan 
scouring pad – do not use any other abrasives.

2.

Fit (new) C-ring 

A

 into recess 

B

.

3.

Locate cell 

C

 and use a small screwdriver to align it with 

test gas injection pipe hole 

D

 in probe body end plate 

E

4.

Carefully slide cell 

C

 into the probe body end plate 

E

5.

Refit test gas injection pipe 

F

 ensuring the short end fits 

into the test gas injection pipe hole 

D

6.

Locate probe filter carrier 

G

 by aligning the internal notch 

H

 with test gas injection pipe 

F

.

7.

Locate 6 M4 x 50 bolts 

I

 and fit nuts 

J

 using an M4 

spanner and 3 mm A/F hexagon wrench – tighten opposing 
bolts evenly to a torque of 4 cNm (5.66 ozf/in).

8.

Locate ceramic filter adaptor 

K

.

9.

Locate the 3 off M4 x 50 bolts 

L

 and fit nuts 

M

 using an 

M4 spanner and 3 mm A/F hexagon wrench – tighten bolts 
evenly to a torque of 4 cNm (5.66 ozf/in).

10. Locate filter gaskets 

N

, ceramic filter 

O

 and clamp plate 

P

.

11. Pass 2 M4 x 60 screws 

Q

 through clamp plate 

P

 and 

fit 2 nuts 

R

 into the recess in ceramic filter adaptor 

K

Tighten bolts evenly to a torque of 4 cNm (5.66 ozf/in) using 
a 3 mm A/F hexagon wrench.

12. Re-install the probe as detailed in probe Instruction manual 

IM/AZ20P-EN

.

13. Make gas supplies to the probe and electrical supplies to 

the transmitter – refer to transmitter Instruction manual 

IM/AZ20E-EN

 and probe Instruction manual 

IM/AZ20P-EN

.

14. Use the new cell’s Zero and Cal. Factor data on the 

Commissioning Label to commission and calibrate the 
probe – refer to transmitter Instruction manual 

IM/AZ20E-EN

 for calibration details.

IMPORTANT (NOTE)

Do not rotate the cell once it has engaged with the 
helical contact at the end of the thermocouple / 
electrode assembly (contained within the probe 
body).

IMPORTANT NOTE

A small amount of anti-seize grease or oil can be 
used on these nuts only to assist assembly. Do not 
use anti-seize grease or oil on any other probe 
fixings.

IMPORTANT NOTE

A small amount of anti-seize grease or oil can be 
used on these nuts only to assist assembly. Do not 
use anti-seize grease or oil on any other probe 
fixings.

IMPORTANT NOTE

Fit the same number of gaskets each side of ceramic 
filter 

O

.

Содержание AZ200 737

Страница 1: ...e AZ20 probe These procedures must be carried out by suitably trained personnel and in accordance with the information given 2 Tools Items required M4 open ended spanner 3 mm A F hexagon wrench Solvent lighter fluid petroleum ether or alcohol surgical spirits Non metallic pan scouring pad Scalpel Small sharp blade Small flat bladed screwdriver Transmitter Instruction manual IM AZ20E EN Probe Instr...

Страница 2: ...sor in a safe manner and in accordance with relevant local regulations Appropriate PPE including mask and goggles must be worn when preparing the process for these procedures WARNING Bodily injury Senor integral transmitter and transmitter weights are detailed in probe Instruction manual IM AZ20P EN Suitable lifting equipment must be available when installing removing the AZ20 components at the pr...

Страница 3: ... the air supply and test gas connections at the sensor 2 Remove the sensor from the process in the reverse order of installation Ensure the sensor cable is free of cable ties restrictions before taking the sensor away from the process DANGER Serious damage to health risk to life Do not proceed until the transmitter is isolated from mains power supplies DANGER Serious damage to health risk to life ...

Страница 4: ...rmocouple electrode assembly carefully withdraw the cell and thermocouple electrode contact assembly until the cell tip G is visible If the assembly will not withdraw sufficiently to access the joint between the cell and helical contact disconnect the red Cell green TC and white TC wires at the probe head end from the inner terminal block connections and straighten the wires refer to probe Mainten...

Страница 5: ...Endura AZ20 series probe Combustion oxygen monitor IN ANAINST 032 EN Rev B 5 Fig 5 1 Removing the diffuser flame arrestor and cell 1 0 t o 1 5 m a x ...

Страница 6: ...cate filter gaskets N ceramic filter O and clamp plate P 11 Pass 2 M4 x 60 screws Q through clamp plate P and fit 2 nuts R into the recess in ceramic filter adaptor K Tighten bolts evenly to a torque of 4 cNm 5 66 ozf in using a 3 mm A F hexagon wrench 12 Re install the probe as detailed in probe Instruction manual IM AZ20P EN 13 Make gas supplies to the probe and electrical supplies to the transm...

Страница 7: ...Endura AZ20 series probe Combustion oxygen monitor IN ANAINST 032 EN Rev B 7 Fig 6 1 Refitting the cell and fitting the large surface area filter ...

Страница 8: ...chnical changes or modify the contents of this document without prior notice With regard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document We reserve all rights in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to thi...

Отзывы: