AW T 210

  H A R T  2- W I R E  T R A N SM I T T ER  |  CO M/AW T 2 10/H A R T- EN  R E V.  B

5

2  Overview

The AWT210 HART® transmitter provides either a 4 to 20mA 
analog signal (conventional analog), a 4 to 20mA analog signal 
with superimposed digital signal (HART point to point analog), 
or a polled digital signal (HART multi-drop).  Using HART 
communications, many functions normally requiring the use of 
the local user-interface can be accessed remotely using HART 
communication devices.

Four variants of AWT210 are available for use with pH, ORP, 
pIon, and conductivity sensors.

The device has four dynamic variables. The device does not 
have any device variables.

The analogue output of this device corresponds to the primary 
variable (pH, ORP, pIon, Ion concentration, conductivity, or 
concentration), outputting 4 mA at the lower range value and 
20 mA at the upper range value. 

Supported Sensors: 

•  pH
•  2-electrode conductivity
•  4-electrode conductivity
•  toroidal conductivity

The ABB AWT210 analyzer complies with HART Protocol 
Revision 7.0.  

Abbreviations and Definitions

Module

Description

TE Module

2-electrode conductivity module

EC Module

4-electrode conductivity module

TC Module

Toroidal conductivity module

Pt100

100 Ω Platinum temperature sensor

Pt1000

1000 Ω Platinum temperature sensor

3k Balco

3000 Ω Balco alloy temperature sensor

ORP

Oxidation-reduction potential

DDL

Device description language

HART

Highway addressable remote transducer

HHT

Hand held terminal

Reference documents

Document ID

Title

DS/AWT210-EN

AWT210 transmitter Data Sheet

OI/AWT210-EN

AWT210 transmitter Operating Instruction

CI/AWT210-EN

AWT210 transmitter Commissioning Instruction

COM/AWT210/
HART-FDS-EN

AWT210 transmitter HART  
Field Device Specification

HART manufacturing ID and device type

All HART products contain unique identifiers that specify the 
product manufacturer and device type.  In accordance with this 
requirement, the ABB Manufacture ID is 26 (1A hexadecimal) 
and the device type code for AWT210 is 34 (22 hexadecimal).

Engineering units and dynamic variables

Dynamic variables

            pH (pH)

   Conductivity (TE, EC, TC)

Measurement

Units

Measurement

Units

PV

pH  
ORP  
pION  
Ion Conc. 

pH  
mV  
%, ppb, ppm,  µg/l  
mg/l

Conductivity 
concentration

µS/cm,  
mS/cm  

%  

ppb 

ppm  

µg/l  

mg/l

SV Temperature

°C, °F

Temperature

°C  

°F

TV

Reference 
impedance

kΩ

Compensated 
conductivity

µS/cm

QV Input voltage

mV

Uncompensated 
conductivity

µS/cm

Unit codes

Unit 

Unit  
code

Unit type/dynamic variable 
classification

Unit type  

code

Deg C

32

Temperature

64

Deg F

33

Temperature

64

uS/Cm

67

Conductivity/Volumetric conductance

87

mS/Cm

66

Conductivity/Volumetric conductance

87

mV

36

EMF

83

pH

59

Analytical

81

Kilo Ohm

163

Electric resistance

85

%

57

Concentration

90

ppm

139

Concentration

90

ppb

169

Concentration

90

ug/l

146

Concentration

90

mg/l

240

Concentration

90

Содержание AWT210

Страница 1: ...al information on installation operation and maintenance For more information Further publications for the AWT210 transmitter are available for free download from www abb com measurement or by scannin...

Страница 2: ...Transmitter functionality and operator interface controls 6 Process interface 6 Sensor Input Channels 6 Host interface 6 Analog output 6 Local interfaces jumpers and switches 6 Local controls and dis...

Страница 3: ...RNING BODILY INJURY Installation operation maintenance and servicing must be performed by suitably trained personnel only in accordance with the information provided in this manual in accordance with...

Страница 4: ...or disposal at no charge to the user ABB is committed to ensuring that the risk of any environmental damage or pollution caused by any of its products is minimized as far as possible IMPORTANT NOTE Fo...

Страница 5: ...re sensor ORP Oxidation reduction potential DDL Device description language HART Highway addressable remote transducer HHT Hand held terminal Reference documents Document ID Title DS AWT210 EN AWT210...

Страница 6: ...tory settings If powered up with SW1 in the OFF position the device retains previously saved user settings SW2 Write Protection If SW2 is in the ON position HW write protect shows as enabled and the u...

Страница 7: ...022 amps x R ohms Load resistance must include any meters external to the AWT210 transmitter the wiring and the system input IMPORTANT NOTE The equation for minimum supply voltage is based on a maximu...

Страница 8: ...tters on the bus Each transmitter on the bus is wired from the primary communication device to the positive and negative POWER SIGNAL terminals of the AWT210 Connect all transmitters on the bus in par...

Страница 9: ...by enabling PDM compatibility at the following menu Device Setup PDM Compatibility Enable Device specific HART commands general Command Description 122 Login to Service Level 123 Read Board Object 124...

Страница 10: ...Parameters 164 Read Factory Calibration Slope and Offset 252 Read PV Sensor Config Table 4 Device specific HART commands Two Electrode conductivity TE Device specific HART commands Toroidal Conductiv...

Страница 11: ...figuration 220 Read Temperature Sensor and Compensation Configuration 221 Write pH Solution Coefficient Value 222 Write millivolt Solution Coefficient Value 223 Read Primary Variable Transfer Function...

Страница 12: ...to make technical changes or modify the contents of this document without prior notice With regard to purchase orders the agreed particulars shall prevail ABB AG does not accept any responsibility wh...

Отзывы: