![ABB AW400 Series Скачать руководство пользователя страница 67](http://html.mh-extra.com/html/abb/aw400-series/aw400-series_instruction-manual_5407067.webp)
65
11 APPENDICES
11.1 EC Declaration
EC Declaration
AW400 line (T17M*4000A) fulfils the compliance to the mandatory tests specified by the applicable harmonised generic rules to
apply the CE Mark, according to European Directive 89/336/EEC and subsequent modifications.
As the final environment may be industrial or residential, the test were performed according to the worst case test set (industrial test
for immunity, residential test for emission). For CE Mark, owing to both the residential and industrial environment applications, the
reference rules are the following “Generic Standards”:
• CENELEC
EN 50081-1/1992
(CEI 110-7) Electromagnetic compatibility - Generic emission standard - Part 1: residential,
commercial and light industry
• CENELEC
EN 50082-1/1992
(CEI 110-8) Electromagnetic compatibility - Generic immunity standard - Part 1: residential,
commercial and light industry
• CENELEC
EN 50081-2/1993
(CEI 110-13) Electromagnetic compatibility - Generic emission standard - Part 2: industrial
environment
• CENELEC
EN 50082-2/1995
(CEI 110-25) Electromagnetic compatibility - Generic immunity standard - Part 2: industrial
environment
TEST DESCRIPTION
LEVELS
EN 55022
Conducted emission
Class B
EN 55022
Radiated emission
Class B
EN 61000-4-4
Electromagnetic compatibility - Part 4: Testing
and measurement techniques - Section 4:
Electrical fast transient/burst immunity test
CM=2kVp on power cable;
CM=2kVp on I/O analog/digital cable;
CM=1kVp on serial cable
EN 61000-4-5
Electromagnetic compatibility -Part 4:
Testing and measurement techniques -
Section 5: Surge immunity test
CM=4kVp, DM=2kVp on power cable
CM=2kVp, DM=1kVp on I/O analog/digital
cable
EN 61000-4-6
Electromagnetic compatibility - Part 4: Testing
and measurement techniques -
Section 6: Conducted disturbances induced by
radio frequency fields immunity test
CM=mod.AM 1kHz
Range 150kHz-230MHz
EN 61000-4-11
Voltage fluctuations
Vn 10%
EN 61000-4-11
Electromagnetic compatibility - Part 4: Testing and
measurement techniques - Section 11: Voltage dips,
short interruptions and voltage variations immunity test
-100% for 5 s
-30% for 10 ms
-60% for 100 ms
EN 61000-4-3
Electromagnetic compatibility - Part 4:
Testing and measurement techniques -
Section 3: Radiated, radio-frequency
electromagnetic field immunity test
10Vrms/m + mod. AM 1kHz
EN 61000-4-2
Electromagnetic compatibility - Part 4:
Testing and measurement techniques -
Section 2: Electrostatic discharge immunity test
AD=8kVp, CD=4kVp
EN 61000-4-8
Electromagnetic compatibility - Part 4: Testing
and measurement techniques - Section 8: Power
frequency magnetic field immunity test
30A/m
AW400 line (T17M*4000B) also fulfils the compliance to the EN61010-1 rules.
Содержание AW400 Series
Страница 1: ...Instruction Manual IM AW4TX Issue 5 Chlorine Transmitter AW400 Series ...
Страница 69: ...67 Notes ...
Страница 70: ...68 ...