Water Hardness Analyzer
AW101 Testomat
11 Specification
20
IM/AW101
Rev. C
11 Specification
Ranges
Determined by reagent – see Table 1.1 on page 2)
Alarms
Three relay outputs –
Limit values (LV1, LV2) and one fault alarm
Resistive load 4 A
Alarm displays
The following faults are shown on the display and also activate the
Alarm output:
Low water pressure
Function fault optics
Measuring fault analysis
Function fault dosing pump
Function fault outlet to drain
Reagent low level
Measuring fault dirtiness
Measuring fault turbine meter
Measuring range exceeded
Front panel indicators
Programme in operation
Analysis stopped
Lack of reagent
Satisfactory result
Unsatisfactory result
Viewing window
Outputs
Current output
0 to 20 mA or 4 to 20 mA
Max. load 500
EMC
Conformity
EN50081-1, EN5008-2, EN61010-1
Power Supply
Voltage
115 V, 230 V or 24 V AC ±10%, 50/60 Hz
Power consumption
30 VA
Unit Protection
115 V, 230 V: T 0.1 A
24 V:T 1.0 A
Environmental Data
Sample pressure
0.1 to 3 bar (1.5 to 45.5 psi)
Sample temperature
10 to 40 °C (41 to 104 °F) max.
Ambient temperature
10 to 45 °C (41 to 124 °F)
Mechanical Data
Ingress protection
IP65
Dimensions
380 x 459 x 280 mm (15 x 18.8 x 11 in.)
Weight
9 kg (19.8 lbs)
Consumables
Reagent consumption
0.07ml (0.000123 pint) per test
D
S/AW10
0
-EN
Rev.
E