ABB AP300 Скачать руководство пользователя страница 14

14

IM/AP300 Rev. J

| AP300 | Combination pH/Redox (ORP) sensors

4 Calibration

4.1 pH Sensor

When the sensor has been correctly connected and all 
electrical connections have been made to the associated pH 
transmitter, it is ready for calibration by either immersing the 
sensor (using suitably sized beakers) either:

1.

in a calibration solution (buffer) of known pH value for a 
single-point calibration,

or

2.

sequentially in two separate calibration solutions of known 
pH values for a two point calibration.

For sensors already in use:

1.

Remove the electrode from the process or sample.

2.

Wash the visible electrode surface with demineralized 
water.

3.

Proceed as described in the paragraph above.

To have agreement with a measured sample, there may be 
times when a process calibration is necessary.

1.

Perform a buffer calibration.

2.

Ensure that the sensor is returned to the process for at 
least 10 minutes before performing a process calibration.

3.

To minimize solution temperature effects, measure the 
sample at the same temperature as the process.

Refer to the instruction manual for the pH transmitter for full 
details of the calibration procedures.

4.2 Redox (ORP Sensor)

When the sensor has been correctly connected and all 
electrical connections have been made to the associated 
Redox (ORP) transmitter, it is ready for calibrating. Follow the 
calibration procedure in the transmitter instruction manual.

For sensors that are connected to transmitters that do not 
have Redox (ORP) sensor calibration capabilities, it is possible 
to check the response as follows:

1.

Prepare standard 4 and 7 pH buffer solutions. Add one 
gram (heaped spatula) of analar quinhydrone to 100 ml of 
each buffer solution. Let them stand for 30 minutes.

2.

Immerse the sensor in each solution in turn and note the 
mV value when stable.

The values obtained should be within ±15 mV of the values 
below:

Warning.

Before removing a sensor from a flow line,

ensure that all isolating valves have been closed.

Caution.

It is important when buffering to ensure that the

visible surfaces of the electrodes have been cleaned using

demineralized  water.  Also  ensure  when  moving  from  one

buffer solution to the next to wash the electrodes and dry

them carefully using a soft tissue.

pH Buffer

mV

4

+259

7

+82

Содержание AP300

Страница 1: ...s a reliable and cost effective solution for most industrial pH applications It is the perfect starting point for pH or Redox measurements Combining measurement reference and temperature compensation elements in a single unit the AP300 is an economical solution for a wide variety of industrial applications including water food pharmaceutical and chemical Sales Service ...

Страница 2: ...1 AP301 Sensors 8 2 1 2 AP302 3 Sensors 9 2 1 3 AP304 5 Sensors 10 3 Electrical Connections 12 3 1 Sensor Connections 12 3 2 Extension Cables 13 4 Calibration 14 4 1 pH Sensor 14 4 2 Redox ORP Sensor 14 5 Maintenance 15 5 1 General Cleaning 15 5 1 1 General Sludge and Loosely Adhering Matter 15 5 1 2 Heavy Non Greasy Deposits 15 5 1 3 Greasy or Organic Deposits 15 5 2 Fault Finding 15 5 3 Storage ...

Страница 3: ... purpose insertion flow through or immersion style The sensor body is molded from chemically resistant PPS Ryton The sensor can be adapted to 1 inch fittings by a threaded Ryton twist lock adaptor An optional PVC electrode guard protects the electrode in immersion applications Dimensions in mm in Fig 1 1 AP301 Sensors and Dimensions 165 1 6 50 33 0 1 30 120 0 4 73 2 5 0 10 64 5 2 54 24 4 0 96 26 7...

Страница 4: ...ral protection for normal bulb style glasses Dimensions in mm in Fig 1 2 AP302 3 Sensors and Dimensions 28 0 1 1 127 0 5 00 22 2 0 80 95 0 3 74 120 0 4 73 238 1 9 38 71 1 2 80 Rear of Sensor and Cable to be Sealed in Conduit or Pipe Customer Supplied 3 4 in Coupling Customer Supplied AP302 Flush Sensor AP303 Notched Sensor Immersion Applications AP303 Notched Sensor Flow Applications AP303 Notched...

Страница 5: ...a standard 11 4 in full port or 11 2 in ball valve Ease of disassembly aids sensor replacement Connection to the ball valve is by compression fitting which is available in either hand tight with 11 4 in NPT threads or wrench tight with 1 in NPT threads AP304 models have no sensor guard flush AP305 models have a notched sensor guard Fig 1 3 AP304 Ball Valve Dimensions Dimensions in mm in Fig 1 4 AP...

Страница 6: ...ush Sensor 11 4 in NPT Full Port Ball Valve Optional Compression Fitting Hand 316 Stainless Steel 11 4 in NPT Dia 3 4 in Conduit Port 3 4 in NPT PVDF Back Piece Anti Blowout Lip Sensor Body No 019 Viton O Rings 316 Stainless Steel Sheath Replacement AP305 Sensor Detail Rear Extension Electrode Protection Tips Viton O Rings No 019 Viton O Rings Stainless Steel Sheath Sensor Screws into Rear Extensi...

Страница 7: ...the sensor is to be expected e g fibrous coatings Use an in line flush sensor body with flat glass sensor mounted at 90 for optimal self cleaning Fig 2 1 Recommendations for Fouling Applications Note The flow of sample passing the sensor helps to keep the sensor clean Sensors should be positioned such that they are always immersed in the sample Not for fouling applications For fouling applications...

Страница 8: ...us measurement Most of the time this can prevented by rotating the sensor to ensure that it is fully immersed in the sample Immersion Type 1 The use of an immersion guard is recommended to protect the glass sensor 2 Levels in many tanks sumps and channels vary The sensor must be immersed to the lowest representative level to ensure the sensor is always immersed in sample 3 Sensor cables on immersi...

Страница 9: ...it is fully immersed in the sample Immersion Type 1 Levels in many tanks sumps and channels vary The sensor must be immersed to the lowest representative level to ensure it is always immersed in sample 2 Sensor cables on immersion sensors should be of adequate length for the BNC to be attached to an extension cable outside the immersion area Fig 2 3 Installing the AP302 3 5 T C Connector if applic...

Страница 10: ...e due to pressure surges or water hammer Insert the sensor only as far as necessary to achieve representative flow Over insertion may damage the sensor and or cause slow response Fully retract the sensor before closing the valve and fully open the valve before inserting the sensor Fig 2 4 Installing the AP304 5 Ball Valve and Compression Fitting 5 T C Connector if applicable Male BNC Cable Rear of...

Страница 11: ...he sensor mark off total onto the shaft of the sensor 5 Insert the sensor into the ball valve Tighten onto the ball valve on reducer 6 Open the ball valve fully and gently push the sensor into the valve until the mark made at 4 reaches the compression fitting 7 Tighten the outer compression joint manually or using a wrench depending on type of fitting chosen 8 A mark can also be made distance B fr...

Страница 12: ...e compensation if applicable Fig 3 1 Cable Terminations Tagged Sensor Terminations for Direct Connection to Transmitter pH with temperature compensation BNC T C Sensor Terminations for Junction Box Connection to Extension Cable pH with temperature compensation Male BNC Male BNC pH without temperature compensation pH and Redox ORP without temperature compensation Glass Electrode Reference Electrode...

Страница 13: ...nector Female BNC to Sensor or Junction Box pH without temperature compensation Part numbers 5 m 16 ft 1015 160 10 m 32 ft 1015 161 20 m 64 ft 1015 162 30 m 96 ft 1015 163 Tagged Pin Extension Cable 5 core to AX400 Conduit Customer Supplied 3 4 in NPT Junction Box pH Sensor Connections pH Sensor Adaptor 1 in NPT 1 in Flow Cell Part No 7670046 1 in Pipe Flow Application Sample ...

Страница 14: ...s the process Refer to the instruction manual for the pH transmitter for full details of the calibration procedures 4 2 Redox ORP Sensor When the sensor has been correctly connected and all electrical connections have been made to the associated Redox ORP transmitter it is ready for calibrating Follow the calibration procedure in the transmitter instruction manual For sensors that are connected to...

Страница 15: ... the extension cable if used No response to pH buffer or sample 1 Check the sensor has been correctly wired to the transmitter as detailed in Section 3 1 page 12 and the relevant transmitter instruction manual 2 Check the glass sensor membrane is not broken or cracked Unstable readings or drift 1 Check the sensor has been correctly wired to the transmitter as detailed in Section 3 1 page 12 and th...

Страница 16: ...uires cable gland 4TB5023 0162 Cable gland 4TB9515 0244 1 in NPT for adapter 1 in NPT for process connection 7670 046 1 in NPT for adapter 1 in BSPT for process connection 7670 043 1 in BSPT to 1 2 in BSPT polypropylene 7601 420 1 in BSPT to 1 2 in NPT polypropylene 7601 430 Tagged pin extension cables 5 core1015 16X for AX460 and AX466 5 m 16 ft 10 m 32 ft 20 m 64 ft 30 m 96 ft 1015 160 1015 161 ...

Страница 17: ...Reference junction Porous PTFE DS AP300 EN Rev I Standard yellow glass 0 to 14 pH Low temperature blue glass 0 to 10 pH Body 0 to 105 C 32 to 221 F Bulb glass 0 to 105 C 32 to 221 F Flat glass 5 to 100 C 41 to 212 F Blue glass 5 to 60 C 23 to 140 F Redox ORP 0 to 105 C 32 to 221 F pH electrode Glass Redox ORP electrode Platinum Junction PTFE Body AP301 PPS Ryton Body AP302 3 and AP304 5 PVDF Kynar...

Страница 18: ...18 IM AP300 Rev J AP300 Combination pH Redox ORP sensors Notes ...

Страница 19: ...AP300 Combination pH Redox ORP sensors IM AP300 Rev J 19 ...

Страница 20: ...echnical changes or modify the contents of this document without prior notice With regard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document We reserve all rights in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to th...

Отзывы: