Dosing Pump
AAO100
3 Analyzer Settings
6
IM/DOSPMP
Rev. B
3 Analyzer Settings
Access the Wash Cycle Configuration menu – see Section 5 of the AX400 pH analyzer’s User Guide (IM/AX4PH).
Wash Mode
Set to
Cont
. (continuous) for use with the dosing pump.
Wash Frequency
Set the wash frequency required.
Wash frequency is set in 15 minute increments between 15 and 45 minutes, then in 1 hour increments
between 1 and 24 hours.
Wash Duration
Set the wash duration required.
Wash duration is set in 15 second increments between 15 and 45 seconds, then in 1 minute increments
between 1 and 10 minutes.
Recovery Period
Set the recovery period required, between 0.5 and 5.0 minutes in 0.5 minute increments.
Fig. 3.1 Wash Cycle Frequency
Wash Mode
-----
Cont.
Pulsed
Wash Frequency
Hours
Mins
15
Wash Duration
Mins
Secs
15
Recovery Period
Mins
1.0
Wash Duration
Continuous
Recovery Period
t
Frequency