17
Set Up Outputs
Current o/p Span
10.00 mg/l C
Test Current o/p
0.00 mg/l C
A1 Action
Off
Factory
Settings
A1 Setpoint
1.50 mg/l C
A2 Action
Off
Set Up
Parameter
A2 Setpoint
2.50 mg/l C
0 - 10
Current o/p Type
4 to 20 mA
0 - 20
Low
High
Low
High
Default Output
8.00 mg/l C
Factory Settings
Factory Set Code
0000
Diss. Organics
page
Set Up
Outputs
5
PROGRAMMING
These parameters are set at the factory and will not normally require
adjustment on site. However, if the receiver/emitter modules have been
changed then the emitter light will need setting. The full
Factory Settings
page is given in Section 6.4.
5.6
Set Up Outputs
5.7
Factory Settings
Set Up Outputs Page
Set the current output span between the following limits:
Low range sensor: 0 to 2 and 0 to 20 mgl
–1
C
High range sensor: 0 to 20 and 0 to 100 mgl
–1
C
Set the current output to a default in the event of a diagnostic alarm.
Select the current output:
4 to 20mA
0 to 20mA
0 to 10mA
Set the value within the current o/p span.
The instrument automatically transmits a test signal to represent the current
output range.
Alarm 1 Action
This can be set as a high or low alarm contact.
Select
High
or
Low
as appropriate.
Select
Off
to disable this alarm.
Adjust setpoint to a value within the range of the sensor.
Alarm 2 Action
This can be set as a high or low alarm contact.
Select
High
or
Low
as appropriate.
Select
Off
to disable this alarm.
Adjust setpoint to a value within the range of the sensor.
Return to top of
Set Up Outputs Page
.