Electromagnetic Flowmeter Converter
10.
Parameter Setting Overview and Flowmeter Design Options
Meter Location:
TAG-No.:
Primary Type:
Converter Type
Order No.:
Instrument No.:
Order No.:
Instrument No.:
Fluid Temp.:
Voltage Supply:
V
Hz
Liner:
Electrodes:
Excitation Frequency:
Hz
C:
System Zero:
Parameter
Setting Range
Prog. Protection Code
0–255 ( 0 = factory setting)
Language
German, English, French, Finnish, Spanish, Italian,
Dutch, Danish, Swedish
Flowmeter Primary:
10DS2110/3110, DS21, DS41, DS41 > 12”[DN 300], DH,
10DI1425 <20”[DN 500]; 10DI1425 >16”[DN 400], 10DS3111 E > 12”[DN 300]
10D1422, 10D1462/1472
Meter Size:
1/25” - 40” [DN 1 – 1000]
Q
max
:
0.05 Range
max
to 1.5 Range
max
Pulse Factor:
0.01 to 1000 pulses/Eng’g unit
Pulse Width:
0.032 – 2000 ms
Low Flow Cutoff:
0 – 10 % of flow range end value
Damping:
0.125 – 99.99 seconds
Filter:
ON/OFF
Density:
0.01 g/cm
3
– 5.0 g/cm
3
Units Q
max
:
l/s, l/min, l/g,hl/s, hl/min, hl/h, m3/s, m3/min, m3/h, igps, igpm,igph,
mdg, gpm, gph, bbl/s, bbl/min, bbl/h, bls/day, bls/min, bls/h, kg/s,
g/min, kg/h, t/s, t/min, t/h, g/s, g/min, g/h, ml/s, ml/min, ml/h, MI/min,
MI/h, MI/day, lb/s, lb/min, lb/h, uton/min, uton/h, uton/day,
kgal/s, kgal/min, kgal/h
Totalizer Units:
l, hl, m3, igal, gal, mgal, bbl, bls, kg, t, g, ml, MI, lb, uton, kgal
Max. Alarm
%
Min. Alarm
%
Terminals P1/P2:
Alarm MAX
Terminals P3/P4:
Alarm MIN
F/R-Contact 44/45/46
Forward/reverse flow direction signal
Current Output:
0/4–20 mA, 0/2–10 mA, 0–5 mA, 0–10–20 mA, 4–12–20 mA
I
out
at Alarm:
0 %, 3.6 mA Max. I
out
(100 %, 115 %, 130 %)
Detector e. Pipe:
ON/OFF
Alarm e. Pipe:
ON/OFF
I
out
at e. Pipe:
0 %, 3.6 mA Max. I
out
(100 %, 115 %, 130 %)
Threshold:
0–100
Adjust e. Pipe:
Software potentiometer value
1st Display Line:
Q (%), Q (units), Q (mA), Q (m/s), totalizer F/R,
difference totalizer, TAG-Number, bargraph
2nd Display Line:
Q (%), Q (units), Q (mA), Q (m/s), totalizer F/R,
difference totalizer, TAG-Number, blank line, bargraph
1st Line Multiplex:
ON/OFF
2nd Line Multiplex:
ON/OFF
Operating Mode:
Standard/piston pump
Flow Direction:
Forward/reverse, forward
Direction Indication:
Normal, Inverse
Store data in ext. EEPROM
Yes/No
Pulse output:
❏
Active
❏
Passive
Communication:
❏
HART-Protocol
❏
RS 485
❏
Profibus DP
Detector empty pipe:
❏
Yes
❏
No
Alarm:
❏
Yes
❏
No
Ext. zero return:
❏
Yes Ext. totalizer reset
❏
Yes Ext. zero compensation
❏
Yes Ext. zero return