8 Configuration and Electrical Calibration
IM/4690–EN Rev. A
33
8.6.3 Electrical Calibration Page
In this section the actual values denoted by '
xxxxx
' are unimportant and are used only to determine display reading stability when
carrying out the electrical calibration procedure.
Electrical Calibration
Press
to advance to next page
or
Press
to return to the
Operating
page – see Section 6.2.1, page 19.
Calibrate
Select
YES
to access the electrical calibration sequence.
Select
NO
to advance to
Adjust RTX Zero
.
mV Zero1 (0mV)
Set the millivolt source to 0 mV and allow the analyzer display to stabilize.
mV Span1 (3000mV)
Set the millivolt source to 3000 mV and allow the analyzer display to stabilize.
Adjust Retransmission Zero
Set the milliammeter reading to 4 mA.
Adjust Retransmission Span
Set the milliammeter reading to 20 mA.
Note.
The retransmission range selected in the
Set Up Outputs
page does not affect the reading.
Return to the
Operating
page – see Section 6.2.1, page 19.
x x x x x
mV Span1(3000mV)
x x x x x
mV Zero1 (0mV)
Calibrate NO
YES
Adjust RTX Zero
Adjust RTX Span
Yes
ELECTRICAL CAL
Turbidity (FNU)
4 5 17
.
No