27
Alter Sec. Code
0 0 0 0 0
SET UP RETRANS
– – – – –
– – – – –
Alter Cal. Code
0 0 0 0 0
Return to
top of page
Duration Mins
1 0
Frequency Hrs
8
A2 Action High
Low
A2 Setpoint
A2 Type Fail
– – – – –
Temp
pH
Off
Water Wash
A2 Type Redox
Off
Water Wash
Water Wash
Fail
or
Off
Redox
OR
Redox
Redox
pH
T
emp or pH
8 0 0
.
Continued from previous page
7
PROGRAMMING…
…7.4
Set Up Alarms Page
Water Wash (4631/4636 only) /Alarm 2 Action
To select the Water Wash cycle press the
key until
Water
Wash
is displayed.
To select Alarm 2 Type, press the
key until
A2
Type
is displayed, then repeat as
for
Alarm 1 Type
,
previous page.
Wash Frequency
Water wash frequency can be set in 15 minute increments to between 15 and 60
minutes, then in 1 hour increments to between 1 and 24 hours.
Wash Duration
Water wash duration can be set in 15 second increments to between 15 and 60 seconds,
then in 1 minute increments to between 1 and 10 minutes.
Alarm 2 Action
Repeat as for
Alarm 1 Action
.
Alarm 2 Set Point
Repeat as for
Alarm 1 Set Point
.
Alter Secure Parameters Security Code
Set the secure parameters security code to a value between 00000 and 19999.
Alter pH Calibration Security Code
Set the pH calibration security code to a value between 00000 and 19999.
Advance to
Set Up Retransmission Page
.