Siting Requirements
3.1.1
Instruments – Fig. 3.1
Caution.
•
Mount in a location free from excessive vibration.
•
Mount away from harmful vapours and/or dripping fluids.
Information.
It is preferable to mount the
instrument at eye level, allowing an unrestricted view of
the front panel displays and controls
3.1.2
Conductivity Cells – Fig. 3.2
Caution.
Ensure that the integral cable (where
applicable) does not hang against hot or abrasive objects
when the plug is connected to the bulkhead socket.
✶
Note.
Allow sufficient clearance for easy removal of
cell for cleaning – see Section 3.4 for overall dimensions
of cells.
Conductivity
Cell
IP66
C – Within Environmental Limits
55
°
C
Max.
–20
°
C
Min.
B – Within Temperature Limits
Maximum Distance
164 ft. (50m) – cell
constant <0.05
A – Maximum Distance of Instrument to Cell
Acid
– 6% Concentration
C – Acid/Alkali Concentration Limits
Alkali
– 8% Concentration
B – Maximum Pressure Limits
A – Maximum Temperature Limits
Max. Temperature – 230
°
F (110
°
C)
Max. Pressure –
150 p.s.i.
Model
2278
Model
2278
Model
2278
3
MECHANICAL INSTALLATION
Fig. 3.1 Siting Requirements – Instrument
Fig. 3.2 Siting Requirements – Conductivity Cell