Static pressure limits
The transmitter models 266CSX can operate within the
specifications with the following limits:
Sensors
Filling fluid
Static pressure limits
A
Silicone oil
3.5 kPa abs., 35 mbar abs., 0.5 psia
and 0.6 MPa, 6 bar, 87 psi
or 2 MPa, 20 bar, 290 psi
depending on code variant selected*
A Fluorocarbon
(Galden)
17.5 kPa abs., 175 mbar abs., 2.5 psia
and 0.6 MPa, 6 bar, 87 psi
or 2 MPa, 20 bar, 290 psi
depending on code variant selected*
C to N
Silicone oil
3.5 kPa abs., 35 mbar abs., 0.5 psia
and 2 MPa, 20 bar, 290 psi
or 10 MPa, 100 bar, 1450 psi
or 41 MPa, 410 bar, 5,945 psi
depending on code variant selected*
C to N
Fluorocarbon
(Galden)
17.5 kPa abs., 175 mbar abs., 2.5 psia
and 2 MPa, 20 bar, 290 psi
or 10 MPa, 100 bar, 1450 psi
or 41 MPa, 410 bar, 5,945 psi
depending on code variant selected*
* 1 MPa, 10 bar, 145 psi for Kynar-PVDF
Test pressure
The transmitters can withstand a pressure test with the
following line pressure without leaking:
Model Test
pressure
266CSX
1.5 x nominal pressure (static pressure
limit) simultaneously on both sides*
* Meets hydrostatic test requirements of ANSI/ISA-S 82.03.
Temperature limits °C (°F)
Environment
This is the operating temperature.
All models
Ambient temperature limits
Silicone oil
Æ
40 and 85 °C (
Æ
40 and 185 °F)
Fluorocarbon (Galden™)
Æ
40 and 85 °C (
Æ
40 and 185 °F)
All models
Ambient temperature limits
Integrated digital display (LCD)*
Æ
40 and 85 °C (
Æ
40 and 185 °F)
Viton™ gasket
Æ
20 and 85 °C (
Æ
4 and 185 °F)
PTFE gasket
Æ
20 and 85 °C (
Æ
4 and 185 °F)
* If may no longer be possible to read the digital display (LCD) clearly below
Æ
20 °C (
Æ
4 °F) and above 70 °C (158 °F).
Note
For applications in potentially explosive environments, the
temperature specified on the certificate / approval which
depends upon the type of protection sought shall apply.
Process
All models
Process temperature limits
Silicone oil
Æ
40 and 121 °C (
Æ
40 and 250 °F)*
Fluorocarbon (Galden™)
Æ
40 and 121 °C (
Æ
40 and 250 °F)**
Viton™ gasket
Æ
20 and 121 °C (
Æ
4 and 250 °F)
PTFE gasket
Æ
20 and 85 °C (
Æ
4 and 185 °F)
* 85 °C (185 °F) for applications under 10 kPa, 100 mbar abs., 1.45 psia up to
3.5 kPa abs., 35 mbar abs., 0.5 psia
** 85 °C (185 °F) for applications below atmospheric pressure up to
17.5 kPa abs., 175 mbar abs., 2.5 psia