4.3 Trouble-Shooting
Localization of electrical malfunction in the 10DX3111 Magnetic Flowmeter can best be accom-
plished by making static and dynamic measurements on the meter in question and comparing rhe
results with the proper resistance and voltage measurements given in Table 4-1. Only qualified
electronic technicians should be permitted to service the meter.
WARNING
Equipment powered by an AC line voltage presents a potential electric
shock hazard to the user. Make certain that the system power input
leads are disconnected from the operating branch circuit before
attempting electrical interconnections or service.
4.3.1 Magnet Coils
To check the magnetic coils follow this procedure:
1) Power down the converter/driver. Remove the coil wiring from terminals M1 and MR in
the converter/driver. (See converter/driver instruction bulletin for location of terminals.)
2) Measure the resistance; the cable resistance should be as indicated in Table 4-1 (or
more depending on the length of the interconnecting cable).
3) Check for short circuit between either of these wires and terminal 3 on the con-
verter/driver. Continuity indicates a shorted magnetic coil. If a shorted coil is detected, the
meter must be replaced.
4) Reattach coil wiring.
TABLE 4-1. MAGNET COIL RESISTANCE
METER
SIZE (INCHES)
RESISTANCE
IN OHMS
30
22
36
13.2
42
8.4
48
4.8
54
4
60
3
66
*
72
*
78
8
*Consult ABB Instrumentation
10DX3111 INSTRUCTION MANUAL
4-2