5-8
Quality Control and Calibration
For a Code reader:
a. Unplug the reader and plug it in
again.
b. If the reader still doesn't scan cor-
rectly, scan the reset code at upper
right, then scan the programming bar-
code at lower right.
7.
Go to
“Running a Quality Control Sample,”
below.
5.1.4
Running a Quality Control Sample
1.
Be sure that you have already entered control values: see
“Entering Quality Control Values”
.
2.
After accepting the new control values, select
Run QC
.
3.
Mix well by slowly rolling the tube flat between your hands (A), and
then slowly inverting the tube 10-15 times (B).
4.
Select
Run
.
Note:
Before running a QC sample, make sure the control tubes have come to
room temperature. Remove the controls from the refrigerator at least
15 minutes prior to use. Do not use a control that is expired or has
been open for more than 14 days, as it may be degraded.