Filling the Reagent Disc
Sample
fill line
Sample
port
Using the provided 100 µL micropipette.
1
Take a reagent disc out of its pouch.
Discs can be used directly from the refrigerator without warming.
2
Attach a new tip to the end of the pipette.
3
Fill the pipette without air bubbles or air gaps with sample.
4
Tilt the disc to 45° with the sample port above the fill line, so that the
entire sample will flow into the sample chamber.
5
Place the pipette tip into the disc’s sample chamber and release the
sample. Make sure to keep the pipette plunger pressed down until the
pipette tip is removed from the sample port.
6
Discard the pipette tip into a biohazard container.
7
Clean the reagent disc. Use a lint-free tissue to remove any sample
spilled on the outside of the disc, taking care that the tissue does
not withdraw any sample from the sample port. Dispose tissue in a
biohazard container.
Beside the micropipette, any other transfer device can be used to dispense
approximately 100 μL of sample into the disc via the sample chamber.
Note:
When using lithium heparin whole blood tubes, gently invert them
several times before use.