Low Battery
The instrument’s battery level is below 10%. Testing becomes disabled
when the battery level drops below 5%.
Recharge or replace battery as soon as possible.
Touch Instrument Status to view the current battery level.
Alerts 1 of 1
Instrument
Status
Exit Alerts
Screen Components and Their Meanings
(Continued)
i-STAT
Alinity v
— User Manual
650-7100 Rev. A
21
If one or more start up self checks fail, the analyzer will display the Alerts
Screen as shown in the
example
above.
Anatomy of Alerts Screen:
Warning or Lockout Icon:
Indicates status of alert.
Analyzer is locked until requirement is satisfied
or Analyzer warning
Alerts Indicator:
Displays number of alerts
Alert Title
Corrective Action Button:
Displays corrective action pathway
Alert Description:
Displays cause and resolution
Action Buttons:
Displays options for screen navigation
A
B
C
D
E
F
A
B
C
D
E
F