Hardware Guide
Document Number: D1000092206 B
7
pulses and converts the information into a signal. The signal represents the ion
intensity for a particular m/z and the instrument displays this formation as a mass
spectrum. For more information, see
.
Figure 1-2
Ion optics path
The API 5000 system uses the Turbo V™ ion source to produce ions from liquid
samples. The Turbo V source can use either the TIS (TurboIonSpray
®
) probe or the
APCI (atmospheric pressure chemical ionization) probe. The instrument is
configured to perform complex MS/MS analysis, but for less rigorous analytical
requirements it can perform single MS scans.
Parts of the Instrument
This section contains diagrams of the instrument and general descriptions of its
components.
Figure 1-3
Front view
Curtain Gas™ flow
Detector
Exit
Interface Region
Analyzer Region
Curtain
Plate Orifice
IQ1
ST3
Stubbies
Stubbies
Deflector
Q0
ST1
Q1
Q2
ST2
IQ3
IQ2
Q3
QJet
®
Ion Guide
Source Exhaust
Vacuum Chamber
Detector