![AAP LOQTEQ Distal Dorsolateral Humerus Plate... Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/aap/loqteq-distal-dorsolateral-humerus-plate/loqteq-distal-dorsolateral-humerus-plate_manual_3218306011.webp)
aap Implantate AG
Lorenzweg 5
•
12099 Berlin
•
Germany
9
Surgical Technique
Elbow Plating System
Surgical Technique
Distal Humerus Plates
90°
plate placement
180°
plate placement
Preoperative planning
•
Evaluate the fracture type on the basis of
the X-rays / CT and select the appropriate
plate length and screw positions.
Also insert lag screws, if necessary.
N
OTE
:
To prevent extensive diaphyseal stress,
it is recommended that the medial and
lateral plates are not the same length,
with at least a two-hole difference
between the plates.
•
LOQTEQ® Distal Humerus Plates are anato-
mically precontoured to fit to the anatomy
of an average distal humerus.
•
In general, the articular block should be
reduced prior to repositioning the articular
block to the shaft.
•
Bicolumn fractures are common, requiring
fixation via a dual plate construct.
LOQTEQ® Elbow Plating System allows for
90° and 180° plate placement.
Minimum 2 holes difference