Measurements 

 

Measurement with the Desired Systolic Pressure 

 

If your systolic pressure is expected to exceed 230 mmHg, use this 
procedure. 

 

1.  Place the cuff on the arm   

(preferably the left arm).   

 

2. Press 

the 

 START 

 button.  

 

3.  While the zero blinks, press and hold 

the 

 START 

  button until a number 

about 30 to 40 mmHg higher than your 
expected systolic pressure appears.   

 

4.  When the desired number is reached, 

release the 

 START 

  button to start 

measurement. Then continue to 
measure your blood pressure as 
described on the previous page. 

 
 
 
 

Notes for Accurate Measurement 

  Sit down in a comfortable position. Place your arm on a table with your 

palm facing upward and the cuff at the same level as your heart. 

  Relax for about five to ten minutes before taking a measurement. If you 

are excited or depressed by emotional stress, the measurement will reflect 
this stress as a higher (or lower) than normal blood pressure reading and 
the pulse reading will usually be faster than normal. 

  An individual's blood pressure varies constantly, depending on what you 

are doing and what you have eaten. What you drink can have a very strong 
and rapid effect on your blood pressure. 

  This device bases its measurements on the heartbeat. If you have a very 

weak or irregular heartbeat, the device may have difficulty determining your 
blood pressure. 

  Should the device detect a condition that is abnormal, it will stop the 

measurement and display an error symbol. See page 5 for the description 
of symbols. 

  This blood pressure monitor is intended for use by adults only. Consult with 

your physician before using this device on a child. A child should not use 
this device unattended. 

 

Refer to the previous page 

for measurement 

Press and hold

At heart level

Press

Inflation in progress

Release

Desired pressure

 

The last data   

at initial display

 

Содержание UA-767PBT-Ci

Страница 1: ...od Pressure Monitor Model UA 767 BT Ci Instruction Manual Original Manuel d instructions Traduction Manual de Instrucciones Traducci n Manuale di Istruzioni Traduzione Bedienungsanleitung bersetzung 1...

Страница 2: ...he Arm Cuff 12 How to Take Accurate Measurements 12 Measurement 12 After Measurement 12 Measurements 13 Normal Measurement 13 Measurement with the Desired Systolic Pressure 14 Notes for Accurate Measu...

Страница 3: ...e device complies with part 15 of the FCC rules and contains the FCC ID POOWML C40 Compliance with Industry Canada IC 4250A WMLC40 The device is a Continua certified Bluetooth wireless technology enab...

Страница 4: ...ing the device repeatedly If you have had a mastectomy please consult a doctor before using the device Do not let children use the device by themselves and do not use the device in a place within the...

Страница 5: ...ure Diastolic Pressure Pulse Rate Battery Indicator I H B Symbol Irregular heartbeat symbol Heart Mark Battery Cover 1 5V Batteries LR6 R6P or AA Battery Compartment Display Air Connector Plug Air Soc...

Страница 6: ...asurement LOW BATTERY The indicator blinks when battery power is low Replace all batteries with new ones when the indicator blinks Unstable blood pressure due to movement during the measurement Try th...

Страница 7: ...EC directive medical device label R TTE directive transmitter module label Manufacturer REP EC EU representative Date of manufacture WEEE label Refer to instruction manual booklet SN Serial number Po...

Страница 8: ...temperatures Generally four new LR6 batteries will last approximately for six months when used twice for measurement each day Remove the batteries if the device is not to be used for a long time The...

Страница 9: ...result in reduced communication speeds or which may prevent connection In this case switch off the power supply to the device that is not being used or use the monitor in a different location If the m...

Страница 10: ...he new device to be paired Some applications may allow pairing with only one device Follow the steps below to pair or unpair the monitor with a Bluetooth compatible receiver device Also refer to the m...

Страница 11: ...environment so be sure to check that the distance is short enough for a connection to be made after measurement is complete Memory In cases when the receiver device cannot receive measurement data th...

Страница 12: ...6 cm to 24 cm Small cuff UA 279 Arm size The circumference of the biceps Symbols that are printed on the cuff Symbols Function Meaning Recommended Action Artery position mark Set the mark on the arter...

Страница 13: ...ir Rest your arm on the table Do not cross your legs Keep your feet on the floor and straighten your back Relax for about five to ten minutes before measurement Place the center of the cuff at the sam...

Страница 14: ...If you wish to stop inflation at any time press the START button again 3 When inflation is complete deflation starts automatically and the heart mark blinks indicating that the measurement is in progr...

Страница 15: ...tional stress the measurement will reflect this stress as a higher or lower than normal blood pressure reading and the pulse reading will usually be faster than normal An individual s blood pressure v...

Страница 16: ...varies by 25 from the average of all heartbeats during the blood pressure measurement It is important that you be relaxed remain still and do not talk during measurements Note We recommend contacting...

Страница 17: ...ding stress and with medication under a doctor s supervision To prevent hypertension or keep it under control Do not smoke Exercise regularly Reduce salt and fat intake Have regular physical checkups...

Страница 18: ...y power is low LOW BATTERY mark blinks If the batteries are drained completely the mark does not appear Replace all batteries with new ones The cuff is not applied properly Apply the cuff correctly Yo...

Страница 19: ...measurement Measurement range Pressure 20 280 mmHg Pulse 40 200 beats minute Measurement accuracy Pressure 3 mmHg Pulse 5 Power supply 4 x 1 5V batteries LR6 R6P or AA or AC adapter TB 233 Not includ...

Страница 20: ...riodically TB 233C Input 100 240 V Output 6 V 500 mA TB 233BF Input 100 240 V Output 6 V 500 mA TB 233 Input 120 V Output 6 V 500 mA Accessories sold separately Cuff AC adapter Catalog Number Plug TB...

Страница 21: ...on Business Park Abingdon Oxfordshire OX14 1DY United Kingdom Telephone 44 1235 550420 Fax 44 1235 550485 A D ENGINEERING INC 1756 Automation Parkway San Jose California 95131 U S A Telephone 1 408 26...

Отзывы: