manualshive.com logo in svg
background image

 13

 

START

Measurements 

 

Measurement with the Desired Systolic Pressure 

 

Model UA-767NFC is designed to detect the pulse and to inflate the cuff 
to a systolic pressure level automatically.   
Use this method when re-inflation occurs repeatedly or when the results are not 
displayed even if the pressure decreases to 20 mmHg or less. 

 

1.  Place the cuff on the arm (preferably the left arm).   

 

2.  Press and hold the 

START

 button until a 

number about 30 to 40 mmHg higher 
than your expected systolic pressure 
appears.  

 

3.  When the desired number is reached, 

release the 

START

 button to start 

measurement. Continue to measure 
your blood pressure as described on 
the previous page. 

 
 
 

Notes for Proper Measurement   

  Sit down in a comfortable position. Place your arm on a table with your 

palm facing upward and the cuff at the same level as your heart. 

 

  Relax for about five to ten minutes before taking a measurement. If you 

are excited or depressed by emotional stress, the measurement will 
reflect this stress as a higher (or lower) than normal blood pressure 
reading and the pulse reading will usually be faster than normal. 

 

  An individual's blood pressure varies constantly, depending on what you 

are doing and what you have eaten. What you drink can have a very 
strong and rapid effect on your blood pressure. 

 

  This device bases its measurements on the heartbeat. If you have a very 

weak or irregular heartbeat, the device may have difficulty determining 
your blood pressure. 

 

  Should the device detect a condition that is abnormal, it will stop the 

measurement and display an error symbol. Refer to page 5 for the 
description of symbols. 

 

  This blood pressure monitor is intended for use by adults only. Consult 

with your physician before using this device on a child. A child should not 
use this device unattended. 

Refer to the previous page 

for measurement 

Press and hold

the button

At heart level 

Release the button 

at the desired 

systolic pressure 

Содержание UA-767NFC

Страница 1: ...Digital Blood Pressure Monitor Model UA 767NFC Instruction Manual 1WMPD4002380...

Страница 2: ...ement 11 After Measurement 11 Measurements 12 Normal Measurement 12 Measurement with the Desired Systolic Pressure 13 Notes for Proper Measurement 13 Near Field Communication NFC Function 14 Memory Da...

Страница 3: ...temperature humidity direct sunlight shock or dust should be avoided Clean the device and cuff with a dry soft cloth or a cloth dampened with water and a neutral detergent Never use alcohol benzene th...

Страница 4: ...MORY Systolic Pressure Diastolic Pressure Pulse Rate Battery Indicator Heart Mark 1 5V Batteries R6P LR6 or AA Clock Display Clock Button I H B Symbol Irregular heartbeat symbol START Button Proper Fi...

Страница 5: ...igned to provide special protection against electrical shocks EC directive medical device label WEEE label Manufacturer EU representative Symbols that appear on the display Symbols Function Meaning Re...

Страница 6: ...mHg of each other The pressure value did not increase during the inflation The cuff is not applied correctly PUL DISPLAY ERROR The pulse is not detected correctly Apply the cuff correctly and take ano...

Страница 7: ...s not display the data stored in memory The memory data can only be read by an NFC enabled device 3 Deleting all Data Stored in Memory Make sure that the monitor displays the clock Press and hold the...

Страница 8: ...ter turning the device off to replace the batteries If LOW BATTERY mark appears even after the batteries are replaced make a blood pressure measurement The device may then recognize the new batteries...

Страница 9: ...press and hold the button to change the value continuously The setting procedure cannot be reversed If the wrong value is set start the setting procedure again After 30 seconds of non operation the se...

Страница 10: ...e date can be set anywhere between the years 2011 and 2055 3 Select the month using the button Press the button to set the current month and move to day selection 4 Select the day using the button Pre...

Страница 11: ...new one Arm Size Recommended Cuff Size Catalog Number 31 cm to 45 cm Large adult cuff UA CUFFBKLA EC 22 cm to 32 cm Adult Cuff UA CUFFBKAU EC 16 cm to 24 cm Small Adult Cuff UA CUFFBKSA EC Arm size Th...

Страница 12: ...d Small Adult Cuff Use Adult Cuff instead of Large Adult Cuff or Small Adult Cuff Lot number How to Take Proper Measurements For the most accurate blood pressure measurement Sit comfortably at a table...

Страница 13: ...nd heart mark blinks indicating that the measurement is in progress Once the pulse is detected the mark blinks with each pulse beat Note If an appropriate pressure is not obtained the device starts to...

Страница 14: ...tes before taking a measurement If you are excited or depressed by emotional stress the measurement will reflect this stress as a higher or lower than normal blood pressure reading and the pulse readi...

Страница 15: ...rformed If the clock of the blood pressure monitor is not correct the clock will be updated at the same time Clock updated Clock display Transmission starts Transmission ends Clock display Measurement...

Страница 16: ...25 from the average of all heartbeats during the blood pressure measurement It is important that you are relaxed remain still and do not talk during measurements Note We recommend contacting your phy...

Страница 17: ...the doctor s readings and provides a more accurate complete blood pressure history WHO Blood Pressure Classification Standards to assess high blood pressure without regard to age have been established...

Страница 18: ...or body during measurement Make sure you remain very still and quiet during measurement The cuff position is not correct Sit comfortably and still Place your arm on a table with your palm facing upwa...

Страница 19: ...tric measurement Measurement range Pressure 0 280 mmHg Pulse 40 180 beats minute Measurement accuracy Pressure 3 mmHg Pulse 5 Power supply 4 x 1 5V batteries R6P LR6 or AA or AC adapter TB 233 Not inc...

Страница 20: ...lephone 44 1235 550420 Fax 44 1235 550485 A D Australasia Pty Ltd 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D ENGINEERING INC 1756 Automation...

Отзывы: