Guidance and manufacturer’s declaration – electromagnetic immunity

 

The A&D unit is intended for use in the electromagnetic environment specified below. The 
customer or the user of the A&D unit should assure that it is used in such an environment.

 

Immunity  
test

 

IEC 60601   
test level

 

Compliance 
level

 

Electromagnetic 
environment – guidance

Electrostatic  
discharge (ESD)   
IEC 61000-4-2

 

± 6 kV   
contact 
± 8 kV   
air

 

± 6 kV   
contact 
± 8 kV   
air

 

Floors should be wood, 
concrete or ceramic tile. 
If floors are covered with 
synthetic material, the   
relative humidity should 
be at least 30%. 

Electrical fast   
transient/burst  
IEC 61000-4-4

 

± 2 kV for power 
supply lines 
± 1 kV 
for input/output   
lines

 

n.a. 

 

Surge  
IEC 61000-4-5

 

± 1 kV   
differential mode 
±2 kV   
common mode

 

± 1 kV   
differential mode 
±2 kV   
common mode

 

Mains power quality 
should be that of a typical 
commercial or hospital 
environment.

 

Voltage dips,short 
interruptions and 
voltage variations   
on power supply   
input lines   
IEC 61000-4-11

 

< 5% U

 

(> 95% dip in U

T

for 0.5 cycle 
40% U

 

(60% dip in U

T

)  

for 5 cycles 
70% U

 

(30% dip in U

T

)  

for 25 cycles 
< 5% U

 

(> 95% dip in U

T

for 5 s

 

< 5% U

 

(> 95% dip in U

T

for 0.5 cycle 
40% U

 

(60% dip in U

T

)  

for 5 cycles 
70% U

 

(30% dip in U

T

)  

for 25 cycles 
< 5% U

 

(> 95% dip in U

T

for 5 s

 

Mains power quality should 
be that of a typical 
commercial or hospital 
environment. If the user of 
the A&D unit requires 
continued operation during 
power mains interruptions, 
it is recommended that the 
A&D unit be powered from 
an uninterruptible power 
supply or a battery.

 

Power frequency   
(50/60 Hz)   
magnetic field   
IEC 61000-4-8

 

3 A/m 

3 A/m 

 

NOTE : U

T

 is the AC mains voltage prior to application of the test level. 

 
 

 

 
 
 
 
 

Содержание UA-767 Plus BT

Страница 1: ...Digital Blood Pressure Monitor Model UA 767 BT Instruction Manual Manuel d instructions Manual de Instrucciones Manuale di Istruzioni Bedienungsanleitung Gebruiksaanwijzing 1WMPD4001357B...

Страница 2: ...Measurements 7 Normal Measurement 7 Measurement With The Desired Systolic Pressure 8 Notes For Proper Measurement 8 What Is An Irregular Heartbeat 9 Pressure Bar Indicator 9 How The Wireless Connecti...

Страница 3: ...f the FCC rules and contains the FCC ID POOWML C40 Compliance with Industry Canada IC 4250A WMLC40 The device is designed for use on adults only not newborns or infants Environment for use The device...

Страница 4: ...lic Pressure Diastolic Pressure Pulse Rate Battery Indicator I H B Indicator Irregular heartbeat indicator Heart Mark Battery Cover 1 5V Batteries R6P or AA Battery Compartment Display Air Connector P...

Страница 5: ...EMORY FullBattery The battery power indicator during the measurement LowBattery The indicator blinks when battery power is low Replace all batteries with new ones when the indicator blinks Unstable bl...

Страница 6: ...eries with new ones Do not mix old and new batteries It may shorten the battery life or cause the device to malfunction LOW BATTERY mark does not appear when the batteries are drained Battery life var...

Страница 7: ...the center of the cuff at the same height as your heart Remain still and keep quiet during measurement Do not measure right after physical exercise or a bath Rest for twenty or thirty minutes before t...

Страница 8: ...ng inflation Note If you wish to stop inflation at any time press the START button again 3 When inflation is complete deflation starts automatically and the heart mark blinks indicating that the measu...

Страница 9: ...xcited or depressed by emotional stress the measurement will reflect this stress as a higher or lower than normal blood pressure reading and the pulse reading will usually be faster than normal An ind...

Страница 10: ...mend contacting your physician if you see this indicator frequently Pressure Bar Indicator The indicator monitors the progress of pressure during measurement How The Wireless Connection Works At the e...

Страница 11: ...under a doctor s supervision To prevent hypertension or keep it under control Do not smoke Exercise regularly Reduce salt and fat intake Have regular physical checkups Maintain proper weight Why Measu...

Страница 12: ...ow LOW BATTERY mark blinks If the batteries are drained completely the mark does not appear Replace all batteries with new ones The cuff is not fastened properly Fasten the cuff correctly You moved yo...

Страница 13: ...uracy Please contact either your authorized dealer or A D for maintenance Technical Data Type UA 767PBT UA 767PBT G Measurement method Oscillometric measurement Measurement range Pressure 20 280 mmHg...

Страница 14: ...voltage power supply network that supplies buildings used for domestic purposes Recommended separation distances between portable and mobile RF communications equipment and the A D unit The A D unit i...

Страница 15: ...F transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the...

Страница 16: ...ential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage var...

Страница 17: ...235 550420 Fax 44 1235 550485 A D Australasia Pty Ltd 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D ENGINEERING INC 1756 Automation Parkway San...

Отзывы: