10.4. Restoring the calibration data to the factory set values
If the gravity acceleration value or calibration data is changed unintentionally, restore
those values to the factory set values, as follows.
1. Refer to “10.1. Calibration mode” to enter the
calibration mode. The gravity acceleration value is
displayed.
2.
Tap
the
UNITS
switch twice to display “
ClrC
”.
3.
Tap
the
TARE
switch to display
“
ClrC
no
” with
“
no
” blinking.
4.
Tap
the
SAMPLE
switch.
“
ClrC
no
” changes to “
ClrC
go
” with “
go
”
blinking.
To cancel the restoring procedure,
tap
the
switch. The display returns to step 2.
5. When “
ClrC
go
” is displayed,
tap
the
TARE
switch. The factory set values are restored
and “
end
” is displayed.
Then, the display returns to “
ClrC
”
To finish the setting procedure,
touch
the
switch or press the
CAL
switch. The scale turns
the power OFF.
Tap
Tap
Tap
Tap
twice