13

 

6-3. Basic weighing operation 

 

1. Press the    ON/OFF    switch to turn the power ON. 

2. Select a weighing unit using the    UNITS    switch. 

3. When the display doesn’t show zero, press the 

  ZERO    switch to set the display to zero. 

4. When using a tare (container), place the 

container on the weighing pan, and press the 

  TARE    switch to set the display to zero. 

5. Place the object to be weighed on the weighing 

pan or in the container, and wait for the STABLE 
indicator to turn on and read the value. 

6. Remove the object from the weighing pan. 

‰

 The  ZERO  switch will zero the scale if the 

weight value is within ±2% of the weighing 
capacity (kg) around the power-on zero point. The 
ZERO

 

indicator turns on. If the weight value 

exceeds ±2% of the weighing capacity (kg),     
the    ZERO    switch does not work. 

‰

 The   TARE   switch will tare the scale and subtract the weight to zero as a tare 

weight when the weight is a plus value. 

‰

 Weighing is possible up to the weighing capacity. When a tare is used, weighing is 

possible up to the weighing capacity less the tare weight value. 

 

Precautions during operation 

‰

 When reading or storing a value, make sure that the STABLE indicator is turned on. 

‰

 Do not press switches with a sharp instrument such as a pencil. 

‰

 Do not apply a shock load to the scale. 

‰

 Do not place a load on the weighing pan that exceeds the weighing capacity. 

‰

 Keep the scale free from foreign objects. 

‰

 Calibrate the scale periodically to keep weighing accuracy. (Refer to “9. CALIBRATION”.) 

 

 

6-4. Weight display resolution   

 

 

 

 

 

The weight display resolution is a ratio of the minimum display to the weighing capacity. 
The SC/SE series has two types of weight display resolution, normal and high.   

 

  Normal: 1/3,000 
  High: 

 1/6,000 

or 

1/7,500 

(depending on the weighing capacity) 

 

The factory setting is the high resolution. Select the resolution that suits your own 
application by the function setting “

r5eo

”. 

 

‰

 For details about the minimum display and the weighing capacity, refer to “13-1. 

Specifications”.  

‰

 In the counting mode, the scale works with the high resolution regardless of the 

weight display resolution selected by the function setting “

r5eo

”. 

NET

ZERO

STABLE

NET

Object to

be weighed

Weighing pan 

Container 

Press 

Содержание SC-30KAM

Страница 1: ...S SC C S SE E S Se er ri ie es s W Wa as sh h D Do ow wn n S Sc ca al le es s SC 30KAM SC 60KAM SC 150KAM SC 60KAL SC 150KAL SE 30KAM SE 60KAM SE 150KAM SE 60KAL SE 150KAL 1WMPD4002054...

Страница 2: ...duced transmitted transcribed or translated into any language in any form by any means without the written permission of A D Company Limited The contents of this manual and the specifications of the i...

Страница 3: ...ting a weighing unit 12 6 3 Basic weighing operation 13 6 4 Weight display resolution 13 7 COUNTING MODE 14 8 COMPARATOR 15 9 CALIBRATION 17 9 1 Gravity acceleration correction 17 9 2 Calibration usin...

Страница 4: ...Alkaline batteries are used is approximately 3 600 hours A large LCD with a character height of 39 mm is used Two weighing platform sizes are available The SC KAL and SE KAL series have a larger weigh...

Страница 5: ...E KAL KAM series Pole support foot Only for the SE KAM series The cable length from the base to the display is approximately 1 5 m S SC C S SE E S Se er ri ie es s W Wa as sh h D Do ow wn n S Sc ca al...

Страница 6: ...4 4 DESCRIPTION OF EACH PART SC KAL KAM series Display Display pole Weighing pan Spirit level Option slot Battery cover Display rear side Leveling foot Display knob...

Страница 7: ...5 S KAL KAM series Display Display pole Weighing pan Leveling foot Spirit level Option slot Battery cover Display knob Display rear side...

Страница 8: ...HIi OK iLOi Turns on when the comparator results are displayed Weighing units lb oz kg and pcs are available A selected unit turns on Battery indicator The battery indicator changes as the battery cap...

Страница 9: ...d to the next step The ZERO and TARE switches work when the weight value is stable The ZERO switch will zero the scale if the weight value is within 2 of the weighing capacity kg around the power on z...

Страница 10: ...ay pole to the base firmly 5 Lay the base on its side Remove the 4 screws 5 in the illustration and remove the pole support foot 6 Pull the excess cable out of the base See 6 in the illustration Take...

Страница 11: ...e the display pole to the base firmly 5 Turn the display knob counterclockwise to remove the display from the display pole 5 in the illustration 6 Pull the excess cable from inside the base out of the...

Страница 12: ...ft or vibration Installing the scale or storing it for an extended period where the temperature and relative humidity are high and where there may be abrupt changes in temperature may cause scale fail...

Страница 13: ...battery compartment as described in step 3 5 Place the battery compartment back into the display using the positioning tab as a guide Secure the catch into the hole with a clicking sound 6 Turn the b...

Страница 14: ...oad beyond this range the scale shows Remove the load from the weighing pan If the scale is to be turned ON with some item loaded performing a zero calibration with the item loaded will allow the scal...

Страница 15: ...p to the weighing capacity less the tare weight value Precautions during operation When reading or storing a value make sure that the STABLE indicator is turned on Do not press switches with a sharp i...

Страница 16: ...at the right side of the number of samples shows zero 5 Place the correct number of samples on the weighing pan or in the container 6 Confirm that the STABLE indicator is turned on Press the UNITS swi...

Страница 17: ...stable data which are more than 4d d minimum display Refer to 13 1 Specifications Also in the counting mode d is equal to the minimum display of kg mode The upper and lower limit values are maintaine...

Страница 18: ...e The minus sign can be selected as a blinking digit after the least significant digit The PRINT switch alternates the minus sign on and off The blinking shows minus and no sign is plus 4 After settin...

Страница 19: ...avity acceleration correction is not required when the scale will be calibrated using a calibration weight at the place where it is to be used Make sure that the scale is in the weighing mode displayi...

Страница 20: ...The calibration weight value is equal to the weighing capacity factory setting If SPAN calibration is not to be performed turn the power OFF to exit from the calibration procedure 6 To calibrate with...

Страница 21: ...d the TARE switch until the gravity acceleration value appears and release the switch 2 Press the UNITS switch twice and display ClrC 3 Press and hold the UNITS switch until ClrCno appears blinking 4...

Страница 22: ...the following switches UNITS To display the next item PRINT To increase the value of the blinking digit by one to change the parameter When the parameter is changed the STABLE indicator turns off 5 Pr...

Страница 23: ...Reply to commands 0 Comparator disabled 1 Compares all data 2 Compares all stable data 3 Compares data of 4d or 4d 4 Compares stable data of 4d or 4d 5 Compares data of 5d Comparator mode Cp 6 Compar...

Страница 24: ...the scale periodically to keep weighing accuracy 12 2 Error codes Overload error Indicates that an object beyond the weighing capacity has been placed on the weighing pan Remove the object from the we...

Страница 25: ...iation 0 005 kg 0 01 kg 0 02 kg 0 01 kg 0 02 kg Linearity 0 01 kg 0 02 kg 0 05 kg 0 02 kg 0 05 kg Sensitivity drift 20 ppm C 5 C to 35 C 41 F to 95 F Display 7 segment LCD character height 39 mm Displ...

Страница 26: ...24 13 2 External dimensions SC 30KAM SC 60KAM SC 150KAM SC 60KAL SC 150KAL Unit mm 60 117 163 380 38 1 173 300 717 220 300 60 158 530 38 1 173 440 123 717 390 300...

Страница 27: ...25 SE 30KAM SE 60KAM SE 150KAM SE 60KAL SE 150KAL Unit mm 60 118 300 202 380 202 722 300 220 40 60 440 128 530 202 300 390 722 40...

Страница 28: ...m s2 Cape Town 9 796 m s2 Rio de Janeiro 9 788 m s2 Chicago 9 803 m s2 Rome 9 803 m s2 Copenhagen 9 815 m s2 San Francisco 9 800 m s2 Cyprus 9 797 m s2 Singapore 9 781 m s2 Djakarta 9 781 m s2 Stockho...

Страница 29: ...27 World map...

Отзывы: