7. COUNTING MODE 

 

Determines a unit weight (the weight of one piece) from the known pieces of samples, 
and calculates how many pieces are on the weighing pan using the unit weight. 
The unit weight is maintained even if the power is turned OFF. 

 
1. Press the    UNITS    switch to select “pcs”. 

(“pcs” = pieces) 

 
 
2. Press and hold the  UNITS  switch to 

enter the sample unit weight storing mode. 
The numerical value on the left indicates 
the number of samples. 

 
 
3. To change the number of samples, press the 

  PRINT    switch. It may be set to 5, l0, 20, 50   

or l00. 

 
 
4. If necessary, place a container on the 

weighing pan, and press the  TARE

1

         

switch. Confirm that the right side of the 
number of samples shows zero.

 

 
 
 
 
5. Place the correct number of samples on 

the weighing pan or in the container. 

 
 
 
 
6. Confirm that the STABLE indicator is turned on. 

Press the  UNITS  switch to calculate and 
store the unit weight. Remove the samples. The 
scale is set to count objects with this unit weight. 

 

‰

 The total weight of samples should be more than 

shown below, regardless of the number of samples. 

Weighing capacity of 30 kg:  0.05 kg 
Weighing capacity of 60 kg:  0.1 kg 
Weighing capacity of 150 kg:  0.2 kg 

If not, the display shows “

lo ut

” and returns to the 

display of step 5. Increase the number of samples 
(step 3) and try again. 

 
7. Place the objects to be counted on the weighing pan. 

ZERO

ZERO

ZERO

STABLE

Confirm

zero

 

Number of

samples 

Samples

Weighing

pan

Container

Press

Press and hold

 

Press

 

Press

 

Press

 

Содержание SC-150KAL

Страница 1: ...S SC C S SE E S Se er ri ie es s W Wa as sh h D Do ow wn n S Sc ca al le es s SC 30KAM SC 60KAM SC 150KAM SC 60KAL SC 150KAL SE 30KAM SE 60KAM SE 150KAM SE 60KAL SE 150KAL 1WMPD4002054 ...

Страница 2: ...oduced transmitted transcribed or translated into any language in any form by any means without the written permission of A D Company Limited The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice ...

Страница 3: ...cting a weighing unit 12 6 3 Basic weighing operation 13 6 4 Weight display resolution 13 7 COUNTING MODE 14 8 COMPARATOR 15 9 CALIBRATION 17 9 1 Gravity acceleration correction 17 9 2 Calibration using a weight 18 9 3 Restoring to the factory setting values 19 10 FUNCTIONS 20 10 1 Setting the parameters 20 10 2 Function list 21 11 OPTIONS 22 12 MAINTENANCE 22 12 1 Notes on maintenance 22 12 2 Err...

Страница 4: ... Alkaline batteries are used is approximately 3 600 hours A large LCD with a character height of 39 mm is used Two weighing platform sizes are available The SC KAL and SE KAL series have a larger weighing pan and the SC KAM and SE KAM series have a smaller weighing pan Select a model that suits your own application Two types of weight display resolution are available normal 1 3 000 and high 1 6 00...

Страница 5: ...SE KAL KAM series Pole support foot Only for the SE KAM series The cable length from the base to the display is approximately 1 5 m S SC C S SE E S Se er ri ie es s W Wa as sh h D Do ow wn n S Sc ca al le es s SC 30KAM SC 60KAM SC 150KAM SC 60KAL SC 150KAL SE 30KAM SE 60KAM SE 150KAM SE 60KAL SE 150KAL Pole support foot ...

Страница 6: ...4 4 DESCRIPTION OF EACH PART SC KAL KAM series Display Display pole Weighing pan Spirit level Option slot Battery cover Display rear side Leveling foot Display knob ...

Страница 7: ...5 SE KAL KAM series Display Display pole Weighing pan Leveling foot Spirit level Option slot Battery cover Display knob Display rear side ...

Страница 8: ...iHIi OK iLOi Turns on when the comparator results are displayed Weighing units lb oz kg and pcs are available A selected unit turns on Battery indicator The battery indicator changes as the battery capacity decreases as shown below New Replace the batteries LCD Weighing units STABLE indicator ZERO indicator ON OFF switch PRINT switch TARE switch HI OK LO indicators NET indicator UNITS switch ZERO ...

Страница 9: ...ed to the next step The ZERO and TARE switches work when the weight value is stable The ZERO switch will zero the scale if the weight value is within 2 of the weighing capacity kg around the power on zero point The ZERO indicator turns on ZERO operation The TARE switch will tare the scale and subtract the weight to zero as a tare weight when the weight is a plus value In this case the ZERO and NET...

Страница 10: ...lay pole to the base firmly 5 Lay the base on its side Remove the 4 screws 5 in the illustration and remove the pole support foot 6 Pull the excess cable out of the base See 6 in the illustration Take care not to damage the cable And make sure the cable has no slack 7 Pull the cable out to the pole side See 7 in the illustration 8 Put the bundle of cable back into the display pole See 8 in the ill...

Страница 11: ...re the display pole to the base firmly 5 Turn the display knob counterclockwise to remove the display from the display pole 5 in the illustration 6 Pull the excess cable from inside the base out of the display pole See 6 in the illustration Take care not to damage the cable And make sure the cable has no slack 7 Put the bundle of cable back into the display pole See 7 in the illustration 8 Place t...

Страница 12: ...aft or vibration Installing the scale or storing it for an extended period where the temperature and relative humidity are high and where there may be abrupt changes in temperature may cause scale failure Install the scale on a solid and level surface Do not install the scale in direct sunlight Do not install the scale near heaters or air conditioners Do not install the scale where flammable or co...

Страница 13: ...e battery compartment as described in step 3 5 Place the battery compartment back into the display using the positioning tab as a guide Secure the catch into the hole with a clicking sound 6 Turn the battery cover clockwise to attach it Make sure that the groove on the cover is correctly engaged and the cover is closed completely Precautions on using the batteries Do not handle the batteries with ...

Страница 14: ...load beyond this range the scale shows Remove the load from the weighing pan If the scale is to be turned ON with some item loaded performing a zero calibration with the item loaded will allow the scale to turn ON that way Refer to 9 CALIBRATION In this case make sure that the total weight of the item loaded at power on and the object to be weighed is within the weighing capacity 2 Press the ON OF...

Страница 15: ...up to the weighing capacity less the tare weight value Precautions during operation When reading or storing a value make sure that the STABLE indicator is turned on Do not press switches with a sharp instrument such as a pencil Do not apply a shock load to the scale Do not place a load on the weighing pan that exceeds the weighing capacity Keep the scale free from foreign objects Calibrate the sca...

Страница 16: ...hat the right side of the number of samples shows zero 5 Place the correct number of samples on the weighing pan or in the container 6 Confirm that the STABLE indicator is turned on Press the UNITS switch to calculate and store the unit weight Remove the samples The scale is set to count objects with this unit weight The total weight of samples should be more than shown below regardless of the num...

Страница 17: ...e stable data which are more than 4d d minimum display Refer to 13 1 Specifications Also in the counting mode d is equal to the minimum display of kg mode The upper and lower limit values are maintained even if the power is turned OFF The upper and lower limit values are common to the weighing and counting mode Ignore the decimal point when setting the upper limit or lower limit value For example ...

Страница 18: ...ne The minus sign can be selected as a blinking digit after the least significant digit The PRINT switch alternates the minus sign on and off The blinking shows minus and no sign is plus 4 After setting all of the digits press the TARE switch Then the upper limit value is stored and the display goes to the lower limit setting mode 5 Set the lower limit value in the same way as the upper limit valu...

Страница 19: ...ravity acceleration correction is not required when the scale will be calibrated using a calibration weight at the place where it is to be used Make sure that the scale is in the weighing mode displaying kg lb oz or pcs 1 Press and hold the TARE switch until the gravity acceleration value in this example 9 7985 appears and release the switch 2 Press and hold the UNITS switch until the value starts...

Страница 20: ... The calibration weight value is equal to the weighing capacity factory setting If SPAN calibration is not to be performed turn the power OFF to exit from the calibration procedure 6 To calibrate with a weight different from the weighing capacity change the displayed value using the following switches UNITS To shift the digit that is blinking unit of measure included to the right PRINT To increase...

Страница 21: ...ld the TARE switch until the gravity acceleration value appears and release the switch 2 Press the UNITS switch twice and display ClrC 3 Press and hold the UNITS switch until ClrCno appears blinking 4 Press the PRINT switch ClrCno changes to ClrCgo 5 When ClrCgo is displayed press the TARE switch The factory setting values are restored and end is displayed To finish the restoring procedure turn th...

Страница 22: ... the following switches UNITS To display the next item PRINT To increase the value of the blinking digit by one to change the parameter When the parameter is changed the STABLE indicator turns off 5 Press the TARE switch to store the setting value After end the scale returns to the weighing mode To restore the function settings to the factory setting values When Clrfno is displayed press the PRINT...

Страница 23: ...1 Reply to commands 0 Comparator disabled 1 Compares all data 2 Compares all stable data 3 Compares data of 4d or 4d 4 Compares stable data of 4d or 4d 5 Compares data of 5d Comparator mode Cp 6 Compares stable data of 5d Comparison conditions d minimum display 0 Weak fast response 1 Normal normal response sStability Response speed Cond 2 Strong slow response 0 Zero tracking function disabled Zero...

Страница 24: ... the scale periodically to keep weighing accuracy 12 2 Error codes Overload error Indicates that an object beyond the weighing capacity has been placed on the weighing pan Remove the object from the weighing pan Underload error Indicates that the weight sensor receives strong force upward Check if the weighing pan is touching anything or if there is anything in the base There is a possibility that...

Страница 25: ...viation 0 005 kg 0 01 kg 0 02 kg 0 01 kg 0 02 kg Linearity 0 01 kg 0 02 kg 0 05 kg 0 02 kg 0 05 kg Sensitivity drift 20 ppm C 5 C to 35 C 41 F to 95 F Display 7 segment LCD character height 39 mm Display update 10 times per second Operating conditions 10 C to 40 C 14 F to 104 F 85 R H or less non condensing Power supply Size D R20P LR20 batteries x 6 Battery operating Approximately 3 600 hours usi...

Страница 26: ...24 13 2 External dimensions SC 30KAM SC 60KAM SC 150KAM SC 60KAL SC 150KAL Unit mm 60 117 163 380 38 1 173 300 717 220 300 60 158 530 38 1 173 440 123 717 390 300 ...

Страница 27: ...25 SE 30KAM SE 60KAM SE 150KAM SE 60KAL SE 150KAL Unit mm 60 118 300 202 380 202 722 300 220 40 60 440 128 530 202 300 390 722 40 ...

Страница 28: ... m s2 Cape Town 9 796 m s2 Rio de Janeiro 9 788 m s2 Chicago 9 803 m s2 Rome 9 803 m s2 Copenhagen 9 815 m s2 San Francisco 9 800 m s2 Cyprus 9 797 m s2 Singapore 9 781 m s2 Djakarta 9 781 m s2 Stockholm 9 818 m s2 Frankfurt 9 810 m s2 Sydney 9 797 m s2 Glasgow 9 816 m s2 Taichung 9 789 m s2 Havana 9 788 m s2 Tainan 9 788 m s2 Helsinki 9 819 m s2 Taipei 9 790 m s2 Kuwait 9 793 m s2 Tokyo 9 798 m s...

Страница 29: ...27 World map ...