Calibration / 

Continued

To calibrate the scale:

1.  If you are already in Setup Menu Mode, scroll to “

F 12”

 and press the 

ZERO

 key. Otherwise,

turn the scale off and set the Calibration Switch on the back panel to the opposite position.
Turn the unit back on.

The message "

C 0

" appears on the display briefly, followed by a value which remains on the

screen. Allow a 20 minute warm-up period for the load cell and other electronics to become
thermally stable.

2. Press 

ZERO

 to zero the value, then press the 

NET/GROSS

 key to save the zero point value.

3.  The display will momentarily prompt "

C 1

" for the span calibration, followed by "

0.00

" with one

digit flashing. Place the test weight on the platform.

4.  Use the four directional keys to adjust the displayed value to the actual test weight value in

pounds. Increase the flashing digit by pressing the 

lb/kg

 key. Decrease the flashing digit by

pressing the 

ZERO

 key. The position of the flashing digit may be changed by pressing the

PRINT

 key or the 

TARE

 key.

5.  After setting the exact value, press the 

NET/GROSS

 key to save the value.

6.  If the calibration was successful, the display will show "

ECAL

" momentarily, then freeze. Exit

the Calibration mode and enter the Normal Operating Mode by positioning the Calibration
Switch back to its original position.

7.  If the calibration was 

not

 successful, one of the error messages below will appear. Take the

indicated action to correct the problem, then perform a new calibration.

"

Err0"

 - The calibration test weight or the adjusted keyed-in weight is larger than full scale.

Change the calibration test weight or check the keyed-in weight.

"Err1" 

- The calibration test weight or the adjusted keyed-in weight is smaller than 1% of full

scale. Change the calibration test weight or check the keyed-in weight

.

"Err2"

 - Check keyed-in weight with the actual weight placed on platform.

To place the unit back into the Normal Operating mode:

1. Toggle the Calibration Switch back to its original position.

2. The display will go through a digit check, then settle into Normal Operating mode. All front

panel keys will now return to their normal mode of operation.

Содержание SB-150

Страница 1: ...S SB B S Se er ri ie es s D Di ig gi it ta al l B Be en nc ch h S Sc ca al le es s Instruction Manual Revision 1 3 April 5 2000 SB 30 SB 70 SB 150 ...

Страница 2: ...d 150 lb 70 kg capaci ties It comes standard with a bright LED screen for easy readout a removable stain less steel platform and an RS 232C serial communication port These scales can be configured to display their divisions in two separate modes non NTEP mode and NTEP mode The scale is shipped from the factory configured in NTEP mode The NTEP mode is supplied for scales which need to comply with H...

Страница 3: ...he rocker switch to the ON 1 position The scale is now ready for use ON OFF SWITCH DC POWER JACK RS 232 PORT REMOTE DISPLAY ID TAG CALIBRATION SWITCH COVER SB 30 70 ONLY CONNECTOR SB 30 70 150 Back Panel The female 9 Pin RS 232 connector is used to interface to a PC Shown at right are the pin assignments for the connector Pin No Pin Name 2 Receive Data 3 Transmit Data 5 Signal Ground 2 3 5 9 pin R...

Страница 4: ...enever the displayed weight is within 0 25 divisions of true zero NET Indicates that the scale is displaying net weight GROSS Indicates that the scale is displaying gross weight TARE Indicates that the tare weight in the system was established by means of the push button method lb and kg Indicates the unit of the displayed weight STABLE This light indicates that the scale is in equilibrium ...

Страница 5: ...on NET GROSS Toggles between Gross and Net weight display if a tare has been established TARE Used to establish the weight on the platform as the tare weight You cannot establish zero or negative gross weight as a tare Also the scale must not be in motion or in overload lb kg Toggles between lb and kg units if enabled in the User A Menu PRINT Sends Print data to serial port if scale is not in moti...

Страница 6: ...ower off to the unit 2 While holding down the lb kg key turn the power back on 3 When the display shows A1 the unit is in User Menu mode and you can release the lb kg key Shown at right are the direc tional and SET key assignments SET USER MENU CHART A1 Baud Rate 1200 2400 4800 9600 A2 Data Bits Parity 8n 7O 7E A3 Transmission Mode C d A4 Serial Data Format A5 Enable lb kg Key 0 1 0 1 2 3 4 5 7n 6...

Страница 7: ...erial Data Format Selects the data format to be transmitted via the serial port to a printer or computer 0 Consolidated Controls Format 1 Toledo 8213 Format 2 A D NCI CAS Format 3 Transcell Technology Format 4 Detecto ASD Format 5 Triner Format 6 Fairbanks 70 2453 4 Format 0 1 2 3 4 5 6 A5 Disable the lb kg Key Allows the lb kg key to be disabled so that an operator cannot acci dentally press the ...

Страница 8: ...ctional and SET key assign ments SET SETUP MENU CHART F1 Display Mode F6 Zero Range F9 Calib Unit 0 1 F11 Display Check F12 Calibration Press ZERO key to begin Press ZERO key to begin 0 1 100 1 9 To place the unit back into the Normal Operating mode 1 With the scale on or off toggle the Calibration Switch back to its original position Turn the scale back on without pressing any keys 2 The display ...

Страница 9: ...ration Places scale into the calibration routine Scrolling down with the ZERO key one level begins the procedure Press ZERO key to begin sequence F13 Factory Reset This sub menu will reset all parameters in the F and A menus to the default settings USE WITH CAUTION Press the ZERO key twice to exe cute Calibration ON OFF SWITCH DC POWER JACK RS 232 PORT REMOTE DISPLAY ID TAG CALIBRATION SWITCH COVE...

Страница 10: ...nged by pressing the PRINT key or the TARE key 5 After setting the exact value press the NET GROSS key to save the value 6 If the calibration was successful the display will show ECAL momentarily then freeze Exit the Calibration mode and enter the Normal Operating Mode by positioning the Calibration Switch back to its original position 7 If the calibration was not successful one of the error messa...

Страница 11: ... x 0 05 lb 0 70 x 0 02 kg CONSTRUCTION Base Steel and ABS Platform Stainless Steel Feet Non skid Hard Rubber DISPLAY 6 Digit 0 6 7 Segment Green LED OVER CAPACITY ANNUNCIATION 102 of Full Scale Capacity OPERATING TEMPERATURE RANGE 32 F to 104 F 0 C to 40 C PLATTER SIZE SB 30 70 10 x 13 SB 150 13 x 13 POWER SOURCE AC Adapter 12VDC 500mA WEIGHT Net Weight 24 3 lb 11 kg Shipping Weight 27 8 lb 12 6 k...

Страница 12: ...elect the proper settings for the following using the scale s settings Baud Rate Data Bits Parity Stop Bits 1 4 Enable the CAPS LOCK key on your PC 5 This test verifies that the scale is transmitting properly If the scale is set to Demand mode pressing the PRINT key on the scale should echo the weight readout on the PC screen If the Continuous mode has been selected on the scale the PC screen will...

Страница 13: ...sary Try re calibrating the scale if this doesn t solve the problem Otherwise possible load cell damage due to overloading LO Warmup Mode The scale has been turned on without the platter installed Install the platter and then press the ZERO key if neces sary Try re calibrating the scale if this doesn t solve the problem Otherwise possible load cell damage due to shock loading Err 0 Calibration Mod...

Страница 14: ...such goods by Seller discloses that the nonconformity actually exists and was not caused by accident misuse neglect alteration improper installation improper or un authorized repair or improper testing 5 Seller will have a reasonable time to repair or replace such goods THESE WARRANTIES EXCLUDE ALL OTHER WARRANTIES EXPRESS OR IMPLIED ORAL OR WRIT TEN INCLUDING WITHOUT LIMITATION WARRANTIES OF MERC...

Отзывы: