2.Pre-Measurement Procedures - 2.2.Turning the Power On/Off
26
1WMPD4004445
Step 4. Turn the Power switch of the product on.
When the power switch on the operation panel of the
product is turned on, the green LED lights up and the power
turns on.
NOTE
Standby current flows to this product when the AC power
cable is connected to the power outlet.
Remove the power cable when the product will not be used
for an extended period of time.
Power switch
2.2.3. Confirming Normal Startup
The monitor screen of the RA3100 is displayed about one minute
after turning the power on.
NOTE
Do not touch the touch panel until the waveform monitor is
displayed. Doing so may lead to erroneous operations.
Immediately after purchase or initializing the main unit,
measurement is turned off for all channels and no waveform is
displayed.
2.2.4. Preparing for More Precise Measurements
Warm up the product for about 60 minutes after turning on the power in order to perform more precise
measurements.
After the warm up is complete, perform "
" of the input modules.
See "
".
This completes the preparations for measurement.
2.2.5. Turning Off the Power
Step 1. When the Power key on the operation panel is pressed
while the power is on, the shut down process starts and the
[Shut down] dialog box indicated below is displayed on the
center of the screen. Tap the OK key to shut down the
product. Tap the Cancel key to continue without turning off
the power.
If the Power key is pressed again while the [Shut down]
dialog box is displayed, the product automatically shuts down.
CAUTION
Make sure to shut down the product to turn off the power, as directly removing the power cable from
the power outlet without shutting down can damage the files in the internal storage.
Содержание RA3100 Omniace
Страница 1: ...RA3100 Omniace Simple Instruction Manual 1WMPD4004445 ...
Страница 73: ...8 Specifications 8 3 Exterior 1WMPD4004445 73 8 3 Exterior 8 3 1 Main Unit Exterior 394 199 410 334 ...
Страница 74: ...8 Specifications 8 3 Exterior 74 1WMPD4004445 MEMO ...
Страница 75: ...Omniace RA3100 Simple Instruction Manual 1WMPD4004445 1st Edition ...