11.3. Function list
Item
Parameter
Description
0
Auto power-off function disabled
1
Turns off after 5 minutes
2
Turns off after 10 minutes
3
Turns off after 15 minutes
4
Turns off after 30 minutes
Auto power-off
function
poff
5
Turns off after 60 minutes
Turns the power OFF
automatically
0
1/3,000
1
1/6,000 or 1/7,500
2
1/15,000 or 1/12,000
Weight display
resolution
re5o
3
1/30,000
Changes the
minimum display
SAMPLE Proceeds to the next unit
TARE
Selects whether a unit is active or
inactive
Weighing unit
Unit
UNIT Proceeds to the next setting
See “6. SELECTING
A WEIGHING UNIT”
0
Zero tracking function disabled
Zero tracking
trc
1
Zero tracking function enabled
Tracks the zero drift
0
Weak stability & quick response
1
2
3
Weighing
stability /
response
speed
Cond
4
Strong stability & slow response
Response =
time from placing an
object on the pan to
turning on the stable
indicator
0
Backlight always off
1
Backlight always on
2
Turns off 5 seconds after stabilizing
3
Turns off 10 seconds after stabilizing
4
Turns off 15 seconds after stabilizing
Backlight
control
l-it
5
Turns off 30 seconds after stabilizing
Sets the timing to
trun off the backlight.
Backlight turns on by
weight change or
switch operation.
0
Dark
1
2
3
Brightness of
backlight
l-i
4
Bright
Adjusts the
brightness
of the backlight
0
Dot
Decimal point
dp
1
Comma
Factory setting