HW-G Instruction Manual Page 49
HW-G OZIM v1a
16.1 Data Format
Bit Format
Format 1
There are four headers for the type of data and weighing condition.
Stable weighing data
S T
Unstable weighing data
U S
Stable counting data
Q T
Out of range (over)
O L
The weighing data consists of 9 characters including decimal point and polarity.
The polarity is always output.
There are three units. The unit is 3 characters.
In case of “out of range”, numbers become all 9
’
s.
The terminator is always output as CR LF.
Data in the unit of kg
Counting mode
Percentage mode
Out of range
Definition of symbols
CR (carriage return)
C
R
LF (line feed)
L
F
0D[H]
0A[H]
Space
20[H]
Содержание HW-G Series