background image

 

HV-WP/HW-WP Series 

Page 69 

 

 

19. Maintenance

19. Maintenance 

 

  Refer to "4. Caution" for use. 
  Refer to "6.1. Display and Symbols " and corresponding mode for displayed error code. 
  Refer to "14. Calibration (Adjusting the Scale)" for precision weighing. 
  Periodically check the accuracy of weighing. Calibrate the scale, if it is moved to 

another location or the environment has changed. 

 
 

19.1.  Check Points Before Calling Maintenance Service 

 

In this situation 

Confirm these items 

Nothing displayed. 
Scale does not turn on. 

 

Is the AC power cord properly connected? 

 

Is the AC power the correct voltage ? 




The scale does not display 
zero at first. 

 

Check around the weighing pan. 

 

Is there anything on the weighing pan. 

 

Perform zero point calibration. 


 ------ 

 is displayed and 

does not proceed. 

 

The weight value is unstable due to drift, vibration or 
other. A breeze or vibration may be affecting the 
measurement.

 

Check around the weighing pan. 

 

Check the connection of load cell cable. 

 

No zero display when the display is turned on. 
Remove anything that is on the pan. 

 

Perform zero point calibration. 



  Cal e 

 is displayed 

 

Calibration error that meaning "Too heavy".  


 -Cal e 

 is displayed 

 

Calibration error that meaning "Too light".  


    e   

 is displayed 

 

Weighing error that meaning "Over load".  


   -e   

 is displayed 

 

Weighing error that meaning "Under load".  


Fixed display 

 

Do you use the "hold function" ? 

 

Turn off scale and turn it on.  


 

19.1.1. Repair 

 

Do not disassemble/ assemble the scale without an authorized service engineer. It may 
cause an electric shock or damage to the scale, etc. In this case, repair is not covered 
under warranty. Contact your local A&D dealer if your scale needs service or repair. 

 
 
 
 

Содержание HV-15KV-WP

Страница 1: ...Waterproof Digital Platform Scale HV 15KV WP HV 60KV WP HV 200KV WP HW 10KV WP HW 60KV WP HW 100KV WP HW 200KV WP 1WMPD4000211B ...

Страница 2: ...ubject to change without notice at any time to improve the product No part of this manual may be photocopied reproduced or translated into another language without the prior written consent of the A D Company Product specifications are subject to change without any obligation on the part of the manufacture ...

Страница 3: ... Basic Operation 17 7 1 Turing the Scale on off and Weighing 17 7 2 Tare And Net Display 18 7 2 1 The Way of Tare Input by Weighing 18 7 2 2 The Way of Digital Input Preset Tare 18 7 3 Weighing Range for the HV WP series 19 7 4 Mode Switch Changing Unit and Mode 20 8 Counting Mode 22 8 1 Storing a Unit Mass 22 8 2 Counting the number of articles 23 9 Percentage Mode 24 9 1 Storing a 100 Mass 24 9 ...

Страница 4: ...4 15 1 The Procedure for Setting Parameters 44 15 2 Parameter List 45 16 RS 232C Serial Interface 48 16 1 Data Format 49 16 2 Stream Mode 51 16 2 1 Preparation and Performance Examples 51 16 3 Command mode 52 16 3 1 Command List 52 16 3 2 Example of Setting Parameters 55 17 Options 56 17 1 Extension cable OP 02 56 17 2 RS 232C Relay output Buzzer OP 03 57 17 2 1 Installing OP 03 58 17 3 RS 422 RS ...

Страница 5: ...e necessary to eliminate the interference FCC Federal Communications Commission in the U S A 1 1 2 Classification of protection provided by enclosures This equipment is designed to comply with the IP Code of IEC 529 The IP 65 code is explained as follows IP International Protection 6 Against ingress of solid foreign objects Dust tight No ingress of dust 5 Against ingress of water with harmful effe...

Страница 6: ...ighings using six figures The comparator function compares the display value with the upper limit value HI lower limit value LO and displays the result The result can be output by a buzzer if option OP 03 is installed The simple batch function or full dribble batch function can be used for filling up to a target mass value The status of a weighing value can be output if option OP 03 or OP 04 is in...

Страница 7: ...ution Do not pull the load cell cable 3mm allen wrench All Accessories Refer to the accessries list on the next page The combination of accessories is according to the scale model Caution Do not pull the load cell cable Instruction manual Main power cord Main power cord Main power cord Main power Please confirm that the local voltage and receptacle type are correct for your scale Display Unit Pan ...

Страница 8: ... A A A A Options List Cable or option name Accessories A OP 02 5m extension loadcell cable Tapping screw M4x10 A OP 03 RS 232C interface Relay output Buzzer Connector JA TCP0586 A OP 04 RS 422 485 interface Relay output Connector TM BLA9 A OP 13 Roller conveyor for HV 200KV WP HW 100KV WP and HW 200KV WP A OP 14 Roller conveyor for HV 60KV WP and HW 60KV WP A AX KO577A 200 RS 232C cable D sub 25 p...

Страница 9: ... apt to be charged with static electricity or where the relative humidity is lower than 45 RH Plastic and isolators are apt to be charged with static electricity Do not use an unstable power source 4 2 Precautions for Operating the Scale Periodically ensure that the weighing value is correct Calibrate the scale before using and after moving it to another location Do not place anything on the weigh...

Страница 10: ...to the bracket using two 3mm Allen screws Step 5 Select the place for installing the scale Also consider 4 Caution on page 7 Step 6 Adjust the level of the base unit by using the Bubble spirit level and Leveling feet Step 7 Ground the scale using the earth terminal Caution Please confirm that the local voltage and the receptacle type are correct for your scale Step 8 Adjust the angle of the indica...

Страница 11: ...unit Disconnect the loadcell cable connector gently Step 3 Remove the ferrite core and cable clamp from the loadcell cable Step 4 Loosen the knobs to remove the display unit Step 5 Remove four 3mm screws from the bottom cover of the bracket for HV 60KV WP HV 200KV WP HW 60KV WP HW 100KV WP HW 200KV WP Step 6 Carefully remove the load cell cable from the pole and the bracket Especially use care wit...

Страница 12: ... Wind the cable through the ferrite core two times Affix the cable to the rear cover using the cable clamp Step 10 Connect the cable to the connector Close the rear cover Step 11 Confirm the accuracy of the scale An example of a scale installation after the pole has been removed and includes the optional Indicator holder Option AX 043005266 Indicator holder Bracket Allen head screw Allen wrench In...

Страница 13: ... to HV 15KV WP HW 10KV WP Procedure A HV 60KV WP HV 200KV WP HW 60KV WP HW 100KV WP HW 200KV WP Procedure B Procedure A HV 15KV WP HW 10KV WP Secure the grounding cable using a M4 screw in the screw hole between the two hexagon bolts on the base unit bottom side Part of Procedure B HV 60KV WP HV 200KV WP HW 60KV WP HW 100KV WP HW 200KV WP Secure the grounding cable using the screw that secures the...

Страница 14: ...rminal Power line terminal Earth terminal PT READY M 15kg 6kg 3kg STABLE NET ZERO HI OK LO Max 3 6 15kg d 1 2 5g RS 232C DIN connector CAP Please Confirm that the local voltage and receptacle type are correct for your scale Knob for angle adjustment Cable clamp for option Weighing condition Inside of rear panel Base Unit Display Unit Knob for angle adjustment Pole Pan Weighing Pan Leveling Foot Bu...

Страница 15: ...e meaning of the mark is as follows ON The weighing value is within the zero band OFF The full dribble batch process is above the zero band Blinking The start or end of the full dribble batch process in not within the zero band A A A A A A The comparator indicator Using the comparator function and comparing a weighing value with the upper and lower limits the result is indicated Using the full dri...

Страница 16: ...arameter of an item ENTER switch Storing new parameters A A A A A Example Preset tare Entering tare with digital input switch Selecting a figure switch Selecting a number ENTER switch Storing a new tare A A A A Example Hold display The hold display is set using f12 of the function table When the value is nearly zero or changes more than 25 30 digits the hold is canceled A A A A Weighing error Chec...

Страница 17: ...is the zero band A A CAP MAX 3 6 15kg d 1 2 5g Description of the weighing unit weighing range and measurable minimum mass Example Displays the weight value by 5 g up to 15 kg Displays the weight value by 2 g up to 6 kg Displays the weight value by 1 g up to 3 kg A A 6 2 Switches Display and Symbols Meaning A Display ON OFF switch Note Standby status when power is connected A A A ZERO switch When ...

Страница 18: ...are and selecting a calibration mass it is used to select a figure A A A A A A A A MODE switch Changing the current unit While setting modes this switch is used for storing a parameter and proceeding to the next step A A A A F switch Full dribble batch function it is used to finish the process Refer to f10 Hold switch Refer to f12 Comparator function selects polarity A A A A A PRINT switch Used to...

Страница 19: ...brate the scale perform it after turning the scale on for 30 minutes warming up Step 6 Press the ZERO switch to display zero with nothing on the weighing pan Step 7 Place an item on the weighing pan gently Step 8 You can read the mass value after the stability mark is displayed Step 9 Remove the item from the weighing pan Step 10 Turn the scale using the ON OFF switch off Memo With the power cord ...

Страница 20: ...sible to put something into the container Wait for the stability mark to be displayed and to read its net display Step 4 Remove all things on the weighing pan 7 2 2 The Way of Digital Input Preset Tare Step 1 Press and hold the SET switch and press the TARE switch Then the blank or stored tare value is displayed A blank display means that the tare value is zero reset value and blinks Step 2 Set th...

Страница 21: ...point mark is displayed it changes to the minimum range automatically Press the ZERO switch to change to the minimum range when there is nothing on the weighing pan and the zero point mark is not displayed due to net display or zero error A A A A A A A A A f2 1 Manual range Press the RANGE switch to expand the range Press the RANGE switch to change to the minimum range when there is nothing on the...

Страница 22: ...Dribble Batch Function f6 9 Accumulation Function f8 1 Comparator Upper limit HI Comparator Lower limit Final value HI Preliminary value Zero band OK LO LO Pound Ounce Percentage Mode Metric kg Counting Mode Non metric units Inactive Comparator f6 and Inactive Accumulation Function f8 0 Metric unit Active Comparator f6 0 f6 7 Active Simple Batch Function or Active Full Dribble Batch Function f6 8 ...

Страница 23: ...owing parameters are stored in the same memory Therefore the functions can not be used at the same time If you use each function it will be necessary to select the function from the function table to set the parameters of HI OK and LO to weigh it using the function Indicator and Output HI OK LO Upper Lower Comparator Function f6 0 7 Upper limit Lower limit Simple Batch Function f6 8 Final value Pr...

Страница 24: ...tep 4 Put a container item on the weighing pan Press the TARE switch Step 5 Put in samples of number selected at step 3 Press the ENTER switch to store it after the satbility mark is displayed Then the count is displayed Caution When the sample is too light and it is not possible to calculate a unit mass the scale displays lo ut and returns to step 3 Increase the number of samples It is necessary ...

Страница 25: ...nit mass Refer to 8 1 Storing a Unit Mass Step 3 Place the container item only on the weighing pan Press the TARE switch Step 4 Put articles in the container item and read the count Step 5 Remove all things from the weighing pan Storing a unit mass STABLE STABLE Zeroing value Pan A container bowl Counting mode Note The pan shape depends on the scale model ...

Страница 26: ...tep 3 With nothing on the weighing pan press the ZERO switch to display zero Step 4 Place the 100 mass on the weighing pan gently Press the ENTER switch to store the 100 mass after the stability mark is displayed Then the percentage is displayed Caution When the sample is too light and it is not possible to calculate a 100 mass the scale displays lo and returns to step 3 Pressing the ENTER switch ...

Страница 27: ...0 Mass Step 3 If a container is needed place the tare item only on the weighing pan and press the TARE switch Step 4 It is now possible to put something on the weighing pan and read the percentage Step 5 Remove all things from the weighing pan Note The pan shape depends on the scale model Zeroing value Percentage Pan A container bowl Storing a 100 mass Percentage mode STABLE STABLE ...

Страница 28: ...switch in the accumulation function is blinking the current function resets The count and accumulated mass value become zero Caution The accumulation function can be used with the first weighing unit accumulated Parameter List and Word Definition The nearly zero is within 5 digits from the zero point in the unit of kg The digit a unit of display is equivalent to the minimum measurable mass The zer...

Страница 29: ...At each finish of the full dribble batch function the scale accumulates the data automatically Use Packaging articles like a powder it is used for recording the bag number and total mass A A A 10 1 Preparation Setting Parameters Step 1 Turn off the display Press the ON OFF switch while the ZERO switch is pressed and held The function table is displayed Step 2 Press the ENTER switch to display an i...

Страница 30: ... Step 6 Press the MODE switch to display the number of articles and total mass with Example 2 This example accumulates the articles that were removed from the weighing pan The function parameter is set to f8 4 Step 1 Enter into the kg mode using the MODE switch Put all articles on the weighing pan and press the TARE switch Step 2 Press the MODE switch to display Step 3 Press the ZERO switch to res...

Страница 31: ... zero is within 5 digits from the zero point in the unit of kg The digit a unit of display and is equivalent to the minimum measurable mass The zero point is the fundamental starting point to weigh anything Function table Meaning and purpose A f6 0 Pressing the SET switch the scale always compares the current display value Press the SET switch again to stop the comparison A A f6 1 The scale always...

Страница 32: ...eliminary value OK use the same memory The upper lower comparator function the simple batch function and the full dribble batch function can not be used at the same time because these parameters use common memory 11 1 Preparation Setting Parameters Step 1 Turn off the display Press the ON OFF switch while the ZERO switch is pressed and held The function table is displayed Step 2 Press the ENTER sw...

Страница 33: ...Step 8 Press the ENTER switch to store the new parameter and display the blinking LO Step 9 Set the lower limit value by using the following switches switch Selecting the number of a figure switch Selecting a figure F switch Selecting the polarity Step 10 Press the ENTER switch to store the new parameters and display the weighing mode If the lower limit value is greater than the upper limit value ...

Страница 34: ...00kg OK is displayed When the current value is greater than 7 000kg HI is displayed Example 2 This example is set as follows Function table f6 4 Pressing the SET switch after the stability mark is displayed the scale compares the current display value with the upper limit value and the lower limit value immediately Upper limit value HI 2 000kg Lower limit value LO 1 000kg Case Pressing the SET swi...

Страница 35: ...OP 03 or OP 04 to use the relay output of the full dribble batch function In case of building up a filling machine with a scale and hopper the performance and processing product of the system are assumed to be as follows The settings are stored in the scale even if the power is removed Refer to Caution on the next page Zero band indicator LO relay output Preliminary value indicator OK relay output...

Страница 36: ...d Word Definition The gross is a total weighing value where the tare value is not subtracted The net is a measurement value with the tare value subtracted from the gross The zero band is the zero detection level The zero point is the fundamental starting point to weigh anything Comparator Function table Meaning and purpose A f6 9 Full dribble batch function A Full dribble batch sub function Functi...

Страница 37: ...an item of the comparator function f6 Step 3 Select a parameter of the full dribble batch function f6 9 with the switch Step 4 Press the ENTER switch to store the new parameter and display an item of the full dribble batch sub function f10 Step 5 Select a parameter of the full dribble batch sub function f10 0 f10 3 with the switch Step 6 Press the ENTER switch and the F switch to display end Step ...

Страница 38: ... preliminary value using the following switches switch Selecting the number of a figure switch Selecting a figure Step 12 Press the ENTER switch to store the new parameter and display the blinking LO of zero band Step 13 Set a zero band which is greater than the tare value using the following switches switch Selecting the number of a figure switch Selecting a figure Step 14 Press the ENTER switch ...

Страница 39: ...elay output of the comparison Install option OP 03 to use the buzzer output of the comparator The settings are stored in the scale even if the power is removed Refer to Caution on the next page Comparison Condition Gross Zero band LO is displayed and output Final Preliminary Net OK is displayed and output Final Net OK HI are displayed and output Parameter List and Word Definition The gross is a to...

Страница 40: ...nction can not be used at the same time because these parameters use common memory 13 1 Preparation Setting Parameters Step 1 Turn off the display Press the ON OFF switch while the ZERO switch is pressed and held The function table is displayed Step 2 Press the ENTER switch to display an item of the comparator function f6 Step 3 Select a parameter of the simple batch function f6 8 with the switch ...

Страница 41: ...d display the blinking LO of zero band Step 11 Set the zero band using the following switches switch Selecting the number of a figure switch Selecting a figure Step 12 Press the ENTER switch to store the new parameter and display the weighing mode 13 2 Operation and Performance Examples Step 1 Select the parameter f6 8 of the function table Step 2 Set the parameters of the final value preliminary ...

Страница 42: ...the weighing pan it is the function that performs adjustment so as to display the zero point mark Comment The zero point is the fundamental starting point to weigh anything influences the performance of scale Span Calibration The function that adjusts the span with a calibrated mass so the scale can accurately weigh anything within the weighing capacity Comment Span means the range of weighing cap...

Страница 43: ... s2 Calcutta 9 788 m s2 Paris 9 809 m s2 Chicago 9 803 m s2 Rio de Janeiro 9 788 m s2 Copenhagen 9 815 m s2 Rome 9 803 m s2 Cyprus 9 797 m s2 San Francisco 9 800 m s2 Djakarta 9 781 m s2 Singapore 9 781 m s2 Frankfurt 9 810 m s2 Stockholm 9 818 m s2 Glasgow 9 816 m s2 Sydney 9 797 m s2 Havana 9 788 m s2 Taiwan 9 788 m s2 Helsinki 9 819 m s2 Taipei 9 790 m s2 Kuwait 9 793 m s2 Tokyo 9 798 m s2 Lisb...

Страница 44: ...l gravity acceleration using the following switches switch Selecting the number of a figure switch Selecting a figure Step 4 Press the ENTER switch to store the new value Then Cal 0 is displayed Step 5 Press the CAL switch again Then the scale returns to the normal weighing mode 14 2 2 Preparation Step 6 Confirm the environmental conditions as follows Maintain a constant temperature and stable pow...

Страница 45: ...Span Calibration Step 11 Set the value of the calibration mass using the following switches This initial value is according to each model switch Selecting the number of a figure switch Selecting a figure Step 12 Place the mass on the weighing pan which was set at step 11 press the ENTER switch while the stable mark is displayed The scale then calculates the span and stores it Step 13 The scale dis...

Страница 46: ...ess the ON OFF switch while the ZERO switch is pressed and held The function table is displayed Step 2 Set parameters for each item using the following switches switch Selecting the parameter of an item switch Selecting a figure F switch Proceeding to Step 3 without storing the parameter ENTER switch Storing a parameter for the current item and proceeding to the next item Step 3 When the end of th...

Страница 47: ...comes nearly zero or a negative value A A A A f5 4 Auto print When the display becomes a stable value without nearly zero the scale outputs the data automatically Next output can be performed after the display becomes nearly zero A A A A Output mode f5 5 At each finish of full dribble batch function the data is output A Selects the comparator mode A f6 0 Pressing the SET switch the scale always co...

Страница 48: ...n function not used A f8 1 The scale accumulates the data if the F switch is pressed when the display becomes a positive stable value without nearly zero Next accumulation can be performed after the display becomes nearly zero or a negative value A A A f8 2 The scale accumulates the data if the F switch is pressed when the display becomes a stable value without nearly zero Next accumulation can be...

Страница 49: ...ion to hold the current display value When the value becomes nearly zero or the weighing value changes more than 25 30 digits hold display is canceled A A A f12 0 Not used A f12 1 The hold function is ON or OFF alternately by the F switch A Hold f12 2 Displaying the stability mark the display is held A Selects readability and response A f13 0 Fast response sensitive value A f13 1 Ú A Averaging f13...

Страница 50: ...llows AX KO577A 200 RS 232C cable D sub 25 pin 2m AX KO1786 200 RS 232C cable D sub 9 pin 2m Transmission system EIA RS 232C Transmission Asynchronous bi directional half duplex Data format Baud rate 2400 4800 9600 bps Data 7 bits Parity 1 bit Even Start bit 1 bit Stop bit 1 bit Code ASCII Terminator CR LF CR 0Dh LF 0Ah Pin connections Pin No Signal name Direction Description A 1 A 2 RXD Input Rec...

Страница 51: ...nd polarity The polarity is always output There are three units The unit is 3 characters In case of out of range numbers become all 9 s The terminator is always output as CR LF Data in the unit of kg Counting mode Percentage mode Out of range Definition of symbols CR carriage return LF line feed 0Dh 0Ah Space 20h S T 0 0 0 1 2 4 0 c k g Header Weighing value Unit Terminator Q T 0 0 0 0 3 0 0 0 c P...

Страница 52: ...ters In case of out of range numbers become all 9 s The terminator is always output as CR LF Caution When the data is longer than 7 characters in the counting precentage mode or accumulation function the overflow is ignored Data in the unit of kg Counting mode Percentage mode Out of range Definition of symbols CR carriage return LF line feed 0Dh 0Ah Space 20h S T 0 0 2 4 0 k g Header Weighing valu...

Страница 53: ...splay Press the ON OFF switch while the ZERO switch is pressed and held The function table is displayed Step 2 Press the ENTER switch to display f4 Baud rate Select a parameter of Baud rate with the switch Step 3 Press the ENTER switch to display f5 Output mode Select Stream mode f5 0 with the switch Step 4 Press the ENTER switch to display f15 Format Select Format 1 f15 0 with the switch Step 5 P...

Страница 54: ...s the MODE switch Template U Command Reply The scale changes mode and unit Zero The current mass value is set to the zero point This is the same as the ZERO switch Template Z Command Reply The mass value becomes zero and the zero point mark is displayed Tare The current mass value is set to zero after placing a tare container receptacle case etc and the net value is displayed This is the same as t...

Страница 55: ...s stored Caution The upper limit value must be greater than the lower limit value Lower limit value A lower limit value is stored The set value is a 5 digit numerical value with a polarity sign and does not contain a decimal point Template LO parameter Command Reply The lower limit value is stored Caution The lower limit value must be less than the upper limit value Accumulation data output The ac...

Страница 56: ... Reply The value is stored Preliminary value The preliminary value is stored The set value is a 5 digit numerical value with a polarity sign and does not contain a decimal point Template S1 parameter Command Reply The value is stored Zero band The zero band is stored The set value is a 5 digit numerical value with a polarity sign and does not contain a decimal point Template S2 parameter Command R...

Страница 57: ...th the switch Step 3 Press the ENTER switch to display f5 Output mode Select Command mode f5 1 with the switch Step 4 Press the ENTER switch to display f15 Format Select Format 1 f15 0 with the switch Step 5 Press the ENTER switch to display end Step 6 Press the ENTER switch to store the new parameters The scale returns to the normal weighing mode Step 7 When the computer transmits a command the s...

Страница 58: ... cable Do not connect more than two extension cables Note The pan shape depends on the scale model Connector board Packing C Connector case Tapping screw M4x10 of accessory 5m extension cable 2m loadcell cable at least Base unit Indicator Indicator holder To be purchased separately as an option AX 043005266 Cap Attachment The loadcell cable from the base unit Connect to the rear connector on the c...

Страница 59: ...pin 2 m AX KO1786 200 RS 232C cable D sub 9 pin 2 m Pin connections Pin No Signal name Direction Description A 1 HI Output Relay output of HI A 2 RXD Input Receive data A 3 TXD Output Transmit data A 4 LO Output Relay output of LO A 5 SG Signal ground RS 232C A 6 OK Output Relay output of OK A 7 DSR Output Data set ready A 8 COM Relay common terminal A Adaptable connector DIN 8pin JA TCP0586 of ac...

Страница 60: ... Remove eight pairs of screws and O rings from the indicator unit Open the rear panel of the unit Step 3 Remove the standard interface board from the indicator unit Install the option board at the same position Step 4 Connect the cable to the J1 connector on the option board Step 5 Connect the DIN plug to the option board Step 6 Close and secure the rear panel using the screws and O rings removed ...

Страница 61: ... bi directional half duplex Data format Baud rate 2400 4800 9600 bps Data 7 bits Parity 1 bit Even Start bit 1 bit Stop bit 1 bit Code ASCII Terminator CR LF CR 0Dh LF 0Ah Address 01 99 Address parameter f11 of the function table Relay Solid state relay Maximum voltage DC50V Maximum current DC100mA Maximum resistance 8 Ω Adaptable connector TM BLA9 of accessory Pin connections Pin No Signal name D...

Страница 62: ...he rear panel of the unit Step 2 Remove the standard interface board from the indicator unit Install the option board at the same position Step 3 Connect the cable to the J3 connector on the option board Step 4 Wire the connector for J1 and connect it to the option board Step 5 Connect a cable from the transformer to the J2 connector Step 6 Set a interface type with the switch Step 7 Close and sec...

Страница 63: ...transmitted If the address is 23 the reply is as follows Normal response Impossible command response Undefined command response Bit format Host computer HV WP HW WP Address 01 SDA SDB RDA RDB Terminator May be built in the host computer SDA SDB RDA RDB TRM SDA SDB RDA RDB TRM SDA SDB RDA RDB TRM SDA SDB RDA RDB TRM SDA SDB RDA RDB TRM SDA SDB RDA RDB TRM HV WP HW WP Address 02 HV WP HW WP Address ...

Страница 64: ...s is the same as the TARE switch Template T Command Reply The current mass value becomes zero and the net mark is displayed Cancel of tare The display value becomes the gross and the net mark is turned off The tare value becomes zero Template CT Command Reply Gross is displayed and the net mark is turned off Preset tare The tare value is set and the net value is displayed The set value is a 5 digi...

Страница 65: ...ata and number The data and number of accumulations become zero Template CA Command Reply The data and number of accumulations become zero Final value The final value is stored The set value is a 5 digit numerical value with a polarity sign and does not contain a decimal point Template S0 parameter Command Reply The value is stored Preliminary value The preliminary value is stored The set value is...

Страница 66: ...ate the scale again if necessary When installing the roller conveyor if not calibrating the scale again the weighing capacity of the scale decreases by the weight of the roller conveyor OP 13 This option can be used with the following models HV 200KV WP HW 100KV WP HW 200KV WP OP 14 This option can be used with the following models HV 60KV WP HW 60KV WP ...

Страница 67: ...can be changed to 10 20 50 100 A Max count number pieces 15 000 12 000 11 000 A Display 7 segment fluorescent display Character height 20mm A Repeatability g 1 2 5 5 10 20 20 50 100 A Linearity g 1 2 5 5 10 20 20 50 100 A Span drift 20ppm C typ 5 C 35 C A Power rating 230V AC 10 15 50 60Hz 20VA Fuse T200mA or 120V AC 10 15 50 60Hz 20VA Fuse T315mA Power source Please confirm that the local voltage...

Страница 68: ...10 000 12 000 10 000 11 000 A Display 7 segment fluorescent display Character height 20mm A Repeatability g 2 10 20 40 A Linearity g 2 10 20 40 A Span drift 20ppm C typ 5 C 35 C A Power rating 230V AC 10 15 50 60Hz 20VA Fuse T200mA or 120V AC 10 15 50 60Hz 20VA Fuse T315mA Power source Please confirm that the local voltage and receptacle type are correct for your scale 50Hz 60Hz A A Ambient temper...

Страница 69: ...HV WP HW WP Series Page 67 18 Specification Dimensions 90 130 55 334 45 330 615 424 786 35 180 60 90 130 437 250 250 466 HV 15KV WP HW 10KV WP HV 60KV WP HW 60KV WP Models Models 45 240 35 180 ...

Страница 70: ...Page 68 HV WP HW WP Series 18 Specification 390 743 530 110 150 70 440 786 45 45 300 HV 200KV WP HW 100KV WP HW 200KV WP Models ...

Страница 71: ... is displayed and does not proceed The weight value is unstable due to drift vibration or other A breeze or vibration may be affecting the measurement Check around the weighing pan Check the connection of load cell cable No zero display when the display is turned on Remove anything that is on the pan Perform zero point calibration A A A Cal e is displayed Calibration error that meaning Too heavy A...

Страница 72: ...Page 70 HV WP HW WP Series MEMO ...

Страница 73: ...HV WP HW WP Series Page 71 MEMO ...

Страница 74: ...Page 72 HV WP HW WP Series MEMO ...

Отзывы: