26
Class
Item and Parameter
Description
0
Key mode
Accepts the PRINT key only when
the display is stable.
1
Auto print mode A
(Reference = zero)
2
Auto print mode B
(Reference = last stable value)
Outputs data when the display is
stable and conditions of
ap-p
,
ap-b
and the reference value are met.
3
Stream mode
Outputs data continuously.
4
Key mode B (Immediately)
Accepts the PRINT key regardless
of the display condition.
5
Key mode C (When stable)
Accepts the PRINT key immediately
when the display is stable, or waits
for the display to be stable when not.
prt
Data output mode
6
Interval output mode
Uses interval output mode.
0
Plus only
Displayed value>Reference
1
Minus only
Displayed value<Reference
ap-p
Auto print polarity
2
Both
Regardless of displayed value
0
10 digits
1
100 digits
ap-b
Auto print difference
2
1000 digits
Difference between reference value
and displayed value
0
Every measurement
1
2 seconds
2
5 seconds
3
10 seconds
4
30 seconds
5
1 minute
6
2 minute
7
5 minute
int
Interval time
8
10 minute
Interval time for the interval output
mode
(With
prt 6
)
0
No pause
pU5e
Data output pause
1
Pause (1.6 seconds)
Selects the data output interval.
0
Not used
at-f
Auto feed
1
Used
Selects whether or not automatic
feed is performed.
0
No output
1
AD-8121 format
info
GLP output
2
General data format
Selects GLP output method.
0
Not used
dout
Data output
ar-d
Zero after output
1
Used
Adjusts zero automatically after data
is output
Factory settings. Digit is a unit of minimum weighing value