Calibration Test Using the Internal Mass (Only for the FZ-CT series)
This function tests the balance weighing accuracy using the internal mass and outputs the
result.
This is available only when the “GLP output (
info
)” parameter is set to “
1
” or “
2
”. (Calibration test
does not perform calibration.)
1 Connect the AC adapter and warm up the balance
at least one hour with nothing on the pan.
2 Refer to “9. FUNCTION TABLEL” to set “GLP
output (
info
)” to “
1
” or “
2
”
3 Press and hold the CAL key until
CC
in
is
displayed, then release the key.
4 The balance measures the zero point.
Prevent vibration and drafts to affect the balance.
5 The measured zero point data is di played.
6 The balance measures the internal mass.
Prevent vibration and drafts to affect the balance.
7 The value of the internal mass is displayed.
The normal range of the value is as follows:
Model
The internal mass
The normal range
FZ-1200CT
200.000 g
FZ-700CT
FZ-500CT
100.000 g
±0.002 g
8 The calibration test report is output using the
RS-232C interface.
9 The balance automatically returns to the weighing mode.