background image

25

11-2. Function list   

Item

Function number 

Description 

Š

f 1- 0 

Auto power-off disabled 

Auto power-off   
function

f 1- 1 

Auto power-off enabled 

Automatically
power off 

Š

f 2- 0 

Normal (1/3,000 class) 

f 2- 1 

High (1/6,000~1/7,500 class)

Display
resolution

f 2- 2 

Higher (1/12,000~1/15,000 class) 

Legal for trade is 

Normal only. 

f 3- 0 

kg

f 3- 1 

g

Š

f 3- 2 

lb

f 3- 3 

oz

Weighing unit   
when powered   
on

f 3- 4 

lb-oz

Factory setting 
may be different 
depending on the 
area to be 
shipped.

Š

f 4- 0 

2400 bps 

f 4- 1 

4800 bps 

Serial interface 
Baud rate 

f 4- 2 

9600 bps 

Optional RS- 
232C/422/485

Š

f 5- 0 

7 bit / Even parity 

f 5- 1 

7 bit / Odd parity 

Serial interface 
Data bit / Parity 

f 5- 2 

8 bit / Non parity 

Optional RS- 
232C/422/485

f 6- 0 

Stream mode 

f 6- 1 

Command mode 

Š

f 6- 2 

Print key mode 

f 6- 3 

Auto-print mode + data 

Serial interface 
Data output 
mode

f 6- 4 

Auto-print mode +/- data 

Optional RS- 
232C/422/485. 
The UFC format 
is applicable to
f6-2 to 4.

f 7- 0 

Upper & Lower limits weight setting 
mode

Š

f 7- 1 

Target weight setting mode with 
HI/LO limit weight 

Comparison 
mode

f 7- 2 

Target weight setting mode with 
HI/LO limit % of the target weight 

Way to set   
parameters. 

f 8- 0 

Comparator disabled 

Š

f 8- 1 

Compares all data 

f 8- 2 

Compares all stable data 

f 8- 3 

Compares data > +4d or < -4d 

f 8- 4 

Compares stable data > +4d or < -4d 

f 8- 5 

Compares data > +4d 

Comparator 
operating
condition

f 8- 6 

Compares stable data > +4d 

Conditions to 
compare. 
d = minimum   
display division

Š

f 9- 0 

Buzzer does not sound. 

f 9- 1 

Buzzer sounds at LO 

f 9- 2 

Buzzer sounds at OK 

f 9- 3 

Buzzer sounds at LO and OK 

f 9- 4 

Buzzer sounds at HI 

f 9- 5 

Buzzer sounds at LO and HI 

f 9- 6 

Buzzer sounds at OK and HI 

Comparator 
buzzer

f 9- 7 

Buzzer sounds at LO, OK and HI 

Buzzer sounds 
according to   
the comparator 
results

f10- 0 

Š

f10- 1 
f10- 2 
f10- 3 

Response /   
Filtering

f10- 4 

Fast / Weak (sensitive) 

Slow / Strong (stable)

Software 
filtering

Š

 Factory setting 

Содержание FS-6Ki

Страница 1: ...Check Weighing Scales FS 6Ki FS 15Ki FS 30Ki 1WMPD4001332...

Страница 2: ...r serious injury CAUTION A potentially hazardous situation which if not avoided may result in minor or moderate injury This is a hazard alert mark Note This manual is subject to change without notice...

Страница 3: ...hing with Preset Tare 11 7 5 To Clear a Tare Weight 11 7 6 Weight display resolution 12 7 7 Simplified operation mode 12 7 8 LCD backlight 12 8 CHECK WEIGHING 13 8 1 Target weight setting mode 13 8 2...

Страница 4: ...necessary to eliminate the interference FCC Federal Communications Commission in the U S A Classification of protection provided by enclosures The equipment is designed to comply with the IP Code of...

Страница 5: ...ighting and analog sweep display of 60 segments with back lighting Scale may be operated by AC power source or an optional SLA sealed lead acid battery Built in comparator with large and bright LED di...

Страница 6: ...to be chraged with static electricity or where the relative humidity is lower than 45 RH Plastic and isolators are apt to be charged with static electricity Do not use an unstable power source 4 2 Pre...

Страница 7: ...ect the place for installing the scale Refer to Cautions for installing the scale below 2 Adjust the level of the base using the spirit level and leveling feet There is an extra foot under the display...

Страница 8: ...6 6 NAMES AND FUNCTIONS Display column Column support foot Spirit level Weighing pan Leveling foot Earth terminal CAL switch cover Cable hook Display pod 1 5 6 9 8 7 3 2 12 11 10 4 4...

Страница 9: ...the results of the weight comparison STABLE ANNUNCIATOR This annunciator will come ON when the weight reading is STABLE ZERO ANNUNCIATOR This annunciator will come ON when the scale is showing the cen...

Страница 10: ...via the 10 key pad SAMPLE 9 Key The SAMPLE key will register the sample weight as the target In the data entry sequence this key displays the number 9 KEY 8 Key The KEY key allows you to enter a targe...

Страница 11: ...1 STORE 4 Key The STORE key is used to store target weight and or HI LO limits In the data entry sequence this key displays the number 4 DISP 3 Key When the DISP key is pressed the analog weight displ...

Страница 12: ...ossible to have the power automatically switched OFF if zero is displayed for approximately 5 minutes See 11 2 Function list and set the function F1 1 7 2 Selecting a weighing unit Press the UNITS key...

Страница 13: ...known tare weight If you hit the wrong key press the C key to return to the display at step 2 and try again 4 Press the ENT key The weight display changes to net weight mode and the NET annunciator c...

Страница 14: ...s mode the following keys are active f14 1 Set the upper lower limits in the setting F14 0 fist and change it to F14 1 Then the target weight only can be set using the SAMPLE key This setting will be...

Страница 15: ...3 Compare when weight is stable or moving and more than 4d or less than 4d F8 4 Compare when weight is stable and more than 4d or less than 4d F8 5 Compare when weight is stable or moving and more th...

Страница 16: ...n 2 Enter the target weight via the 10 key pad If you hit the wrong key press the C key and the old target weight is re displayed Continue data entry again You can place minus value using the key This...

Страница 17: ...eight 3 000 kg HI limit weight 0 050 kg LO limit weight 0 030 kg Then Upper limit 3 000 kg 0 050 kg 3 050 kg Lower limit 3 000 kg 0 030 kg 2 970 kg LO weight 2 970 kg OK 2 970 kg weight 3 050 kg HI 3...

Страница 18: ...Lower limit 3 000 kg 3 000kg x 0 50 2 985 kg 8 2 Upper Lower limits weight setting mode The UPPER LOWER LIMITS WEIGHT SETTING MODE uses an upper limit weight and a lower limit weight The target weigh...

Страница 19: ...ight 2 950 kg OK 2 950 kg weight 3 050 kg HI 3 050 kg weight 8 3 The comparator memory The FS i scale can store up to 100 comparator limits by 2 digit numbers form 00 to 99 To use this function the sc...

Страница 20: ...f you stop to recall press the C key the scale returns to the weighing mode 3 Press the ENT key to recall the data from the memory 4 The display shows a set of data for a few seconds one by one and re...

Страница 21: ...60 segments represent zero to full scale When the scale is in the net mode the analog sweep display shows net weight In the TARGET WEIGHT SETTING MODE the 40th segment will flash as the target weight...

Страница 22: ...15 00 kg As the weight increases the analog segments approach the flashing segment This will provide a simple way of filling a container to a target weight The UPPER LOWER LIMIT WEIGHT SETTING MODE F...

Страница 23: ...og sweep display FS 30Ki d 0 01 kg TARGET WEIGHT SETTING MODE F07 1 or F07 2 Target weight 5 00 kg HI 0 10 kg LO 0 09 kg 00010 20 Within the limits Above the upper limit 00010 00 00009 80 Lower limit...

Страница 24: ...n mode Press the ON OFF key or CAL switch to exit without calibration 3 Press the ENT key then the gravity acceleration value will be displayed It is not necessary to set the gravity acceleration valu...

Страница 25: ...place it should be calibrated using a calibration weight But if a calibration weight is not available the gravity acceleration correction will compensate the scale Change the gravity acceleration val...

Страница 26: ...the 10 ley pad 4 Press the ENT key Then the parameter stored is appears 3 Enter the parameter value using the 10 ley pad If you do not need to change the parameter turn the power off without pressing...

Страница 27: ...to print mode data Optional RS 232C 422 485 The UFC format is applicable to f6 2 to 4 f 7 0 Upper Lower limits weight setting mode f 7 1 Target weight setting mode with HI LO limit weight Comparison m...

Страница 28: ...right Brightness control f16 0 Simple weighing display mode f16 1 Target weighing display mode f16 2 Limit check weighing display mode Analog sweep display when powered on f16 3 Display off f17 0 Alwa...

Страница 29: ...0 005 kg Linearity error 0 002 kg 0 005 kg 0 01 kg Span drift 20 ppm C 5 C 35 C 41 F 95 F Display 7 segment LCD display Character height 18 6 mm with backlight 60 segment analog sweep display with bac...

Страница 30: ...28 12 2 Dimensions FS 6Ki FS 15Ki FS 30Ki Unit mm 250 250 250 98 48 414 496 250 380 464 496 107 300 48...

Страница 31: ...m s2 Cape Town 9 796 m s2 Rio de Janeiro 9 788 m s2 Chicago 9 803 m s2 Rome 9 803 m s2 Copenhagen 9 815 m s2 San Francisco 9 800 m s2 Cyprus 9 797 m s2 Singapore 9 781 m s2 Djakarta 9 781 m s2 Stockh...

Страница 32: ...30 World map...

Страница 33: ...31 MEMO...

Страница 34: ...32 MEMO...

Страница 35: ......

Страница 36: ......

Отзывы: